Key result
Active surveillance of over 370,000 ICD leads using the NCDR registry found no difference in lead survival rates between matched groups, with failure-free survival ranging from 97.7% to 98.9%.
Answers are grounded in this paper and cite their sources.
Active surveillance of clinical registries using novel algorithms offers a promising, though imperfect, paradigm for postmarket monitoring of ICD lead survival.
HomeCirculation: Cardiovascular Quality and OutcomesVol. 13, No. 4Defibrillator Lead Survival Free AccessEditorialPDF/EPUBAboutView PDFView EPUBSections ToolsAdd to favoritesDownload citationsTrack citationsPermissions ShareShare onFacebookTwitterLinked InMendeleyReddit Jump toFree AccessEditorialPDF/EPUBDefibrillator Lead SurvivalWhere Is the Threshold? Emily P. Zeitler, MD, MHS and Kimberly A. Selzman, MD, MPH Emily P. ZeitlerEmily P. Zeitler Emily P. Zeitler, MD, MHS, Cardiac Electrophysiology, Dartmouth-Hitchcock Medical Center, Geisel School of Medicine at Dartmouth, The Dartmouth Institute, 1 Medical Center Dr, Lebanon NH 03756. Email E-mail Address: [email protected] Dartmouth-Hitchcock Medical Center and The Dartmouth Institute, Lebanon, NH (E.P.Z.). Geisel School of Medicine at Dartmouth, Hanover, NH (E.P.Z.). and Kimberly A. SelzmanKimberly A. Selzman George E. Wahlen Department of Veterans Affairs Medical Center and University of Utah School of Medicine, Salt Lake City, UT (K.A.S.). Originally published14 Apr 2020https://doi.org/10.1161/CIRCOUTCOMES.120.006649Circulation: Cardiovascular Quality and Outcomes. 2020;13:e006649This article is a commentary on the followingActive Surveillance of the Implantable Cardioverter-Defibrillator Registry for Defibrillator Lead FailuresOther version(s) of this articleYou are viewing the most recent version of this article. Previous versions: April 14, 2020: Ahead of Print See Article by Resnic et alLife-sustaining medical devices in the United States undergo rigorous evaluations before market approval, and implantable cardioverter-defibrillator (ICD) leads are no exception. For transvenous ICD leads, this typically includes a battery of bench tests and in vivo testing. However, this type of premarket evaluation is limited by our collective experience with prior lead failures and the imagination of the engineers designing the testing. When devices fail in novel, rare, or time-delayed ways, traditional premarket testing may be ill-equipped. Recognition of this led to mandatory 5-year postapproval studies for defibrillator leads starting in 2008. However, these postapproval studies proved to be expensive and resource-intensive, and the results failed to provide meaningful information about lead failures in a timely way. Other postmarket surveillance systems, like the Manufacturers and User Facility Device Experience, are, at best, incomplete since reporting is voluntary.1 Acknowledgement of this has led to an ongoing effort to reevaluate postmarket device surveillance. Regulators, payers, and manufacturers share the common goal of achieving a postmarket surveillance system that is more nimble and timely while being more cost-effective and resource-efficient.In recognition of the shortcomings of existing postmarket surveillance strategies, Resnic et al2 have performed a thoughtful analysis in this issue of Circulation: Cardiovascular Quality and Outcomes. These investigators who are long-time leaders in device surveillance explored whether existing data sources coupled with novel algorithms could serve as a new paradigm for postmarket device surveillance in hopes that a larger but less detailed dataset could efficiently monitor ICD lead survival. The investigators harnessed the National Cardiovascular Data Registry (NCDR) for ICD implantation during a period of near-universal registry participation (2011–2016) to evaluate rates of ICD lead failure for each of the 4 most frequently implanted ICD leads of that period (lead models were masked). The investigators applied Data Extraction and Longitudinal Trend Analysis (DELTA), an open-source software package for active clinical data surveillance, which has been used previously with a variety of data sources, including other NCDR registries.3 For each of the successive 3-month periods over 5 years, DELTA generated 4 matched pairs—one for each lead model and a propensity-matched comparator group from patients receiving all other lead models from the same period. By matching lead implant procedures to any prior lead procedure in the same patient, lead survival was examined. Lead survival was divided into short (<30 days) and late term (>30 days) since early and late failure mechanisms tend to be quite different. The investigators recognized that given the size of this dataset, statistically significant and clinically meaningful differences in lead failure rates may not always align, so a prespecified threshold for a safety alert was selected as twice the rate in the matched control group.Predictably, the cohort was enormous—comprising nearly all ICD implants in Medicare beneficiaries in the US during the 5-year period—with over 370 000 ICD leads analyzed. Furthermore, the propensity score matching was excellent. There was no difference in the lead survival rate between groups with a failure-free survival range of 97.7% to 98.9%. A large proportion of lead failures occurred within 30 days of implant (44%). Importantly, the predefined safety threshold was not met during any 3-month period for any lead model.Resnic et al2 should be applauded for their repurposing of clinical data for safety surveillance in this collaborative effort between NCDR, Food and Drug Administration, and the DELTA analytic center. However, there are some limitations to this approach for ICD lead surveillance. First, the authors use "lead replacement" as a surrogate for lead failure. In general, these outcomes are probably highly correlated and unlikely to vary from one lead model to another, but it is a surrogate nonetheless. Experience shows that true lead failure is best assessed with electrical parameters and imaging, neither of which are available in the NCDR-ICD Registry. This shortcoming reveals one of the problems with repurposing existing data systems: the NCDR-ICD Registry was never intended to operate as a postapproval surveillance mechanism; its design was not fit to purpose for lead survival analysis. One of the fundamental limitations in this regard is the lack of longitudinal follow up data. Resnic et al2 addressed this by matching lead implants with prior procedures at the patient level—a meticulous method that relies on near-universal inclusion. However, even this careful method cannot differentiate between true lead failure and "lead replacement" for other reasons, like dislodgement. Integration of additional lead survival data, such as manufacturer remote monitoring transmissions, with those from the Registry, would create a more robust data source for lead surveillance and improve the specificity of lead failure.Second, the authors compare rates of lead replacement between lead models rather than to a predetermined performance criterion. This approach was taken to cancel out clinically insignificant lead failure noise present in all lead models to identify true lead failure signal. However, in the case of ICD leads, acceptable rates of lead survival are well established, so one could argue that the appropriate comparator is an objective performance criterion. In such a scenario, performance by a single lead model that falls below this criterion could trigger a deeper dive into lead performance, leveraging additional information, such as electrical performance and claims data. Indeed, this is one of the proposed methods for a new paradigm in lead surveillance currently being developed by a consortium of academics, regulators, and industry representatives in the Predictable and SuStainable Implementation of National Cardiovascular Registries for Electrophysiology (EP PASSION).4–6Third, applying the DELTA algorithm to the NCDR data is a proof of concept exercise. It could have been further validated by retrospectively evaluating the survival of a lead with a known elevated failure rate that reached regulatory threshold for a class I recall. For example, why not use this approach to evaluate the survival of the St Jude Medical Riata ICD lead in the period before the recall? The Riata story unfolded squarely within the time period of interest presented here, so a look back may have offered an opportunity to validate this methodology. Lead failure rates are also available in product performance reports, which are industry-generated documents describing device performance based on operator and manufacturer reports along with returned product analysis. This source offers another possible mechanism to validate these findings. Although this data source includes certain biases that likely underestimate failure, it could serve as a rough comparator. Indeed, the product performance reports for the 4 leads explored by Resnic et al2 show a 5-year lead survival of 97% to 99% during the time period of interest.7–10Last, the authors likely did not know, when creating the protocol in 2014, that participation in the NCDR-ICD Registry would cease being a requirement for coverage of primary prevention ICDs for Medicare beneficiaries in February 2018.11 It remains unclear how this update will change participation in the Registry and, thus, whether this Registry will continue to be adequately representative of the ICD population overall.Despite these limitations, this work by Resnic et al2 is a great step towards repurposing clinical data for public health and safety. In our proinnovation world where advancing medical device technology is valued, and postmarket reporting is not universally mandatory, we must have robust and sustainable methods to continue learning about devices after Food and Drug Administration approval. This reality poses challenges to existing data collection and interpretation mechanisms. Novel surveillance approaches are needed and will vary depending on existing data systems, device maturity, payer coverage, and many other factors with a goal for vigilant pairing of systems with purpose. Software solutions, like DELTA, serve as a path forward for repurposing existing data systems to serve the medical device ecosystem and public health at large.DisclosuresNone.FootnotesThe opinions expressed in this article are not necessarily those of the editors or of the American Heart Association.Emily P. Zeitler, MD, MHS, Cardiac Electrophysiology, Dartmouth-Hitchcock Medical Center, Geisel School of Medicine at Dartmouth, The Dartmouth Institute, 1 Medical Center Dr, Lebanon NH 03756. Email emily.p.[email protected]orgReferences1. Resnic FS, Normand SL. Postmarketing surveillance of medical devices–filling in the gaps.N Engl J Med. 2012; 366:875–877. doi: 10.1056/NEJMp1114865CrossrefMedlineGoogle Scholar2. Resnic FS, Majithia A, Dhruva SS, Ssemaganda H, Robbins S, Marinac-Dabic D, Hewitt K, Ohno-Machado L, Reynolds MR, Matheny ME. Active surveillance of the ICD registry for defibrillator lead failures.Circ Cardiovasc Qual Outcomes. 2020; 13:e006105. doi: 10.1161/CIRCOUTCOMES.119.006105LinkGoogle Scholar3. Majithia A, Matheny ME, Paulus JK, Marinac-Dabic D, Robbins S, Ssemaganda H, Hewitt K, Ponirakis A, Loyo-Berrios N, Moussa I, Drozda J, Normand SL, Resnic FS. Comparative safety of aspiration thrombectomy catheters utilizing prospective, active surveillance of the NCDR CathPCI registry.Circ Cardiovasc Qual Outcomes. 2019; 12:e004666. doi: 10.1161/CIRCOUTCOMES.118.004666LinkGoogle Scholar4. National Evaluation System for health Technology Coordinating Center. NEST Coordinating Center Test-Cases.https://nestcc.org/portfolio-item/electrophysiology-predictable-and-sustainable-implementation-of-national-registries-ep-passion/. Accessed March 20, 2020.Google Scholar5. Zeitler EP, Al-Khatib SM, Drozda JP, Kessler LG, Kirtane AJ, Kong DF, Laschinger J, Marinac-Dabic D, Morice MC, Reed T, Sedrakyan A, Stein KM, Tcheng J, Krucoff MW; MDEpiNet. Predictable and SuStainable Implementation of National Cardiovascular Registries (PASSION) infrastructure: a think tank report from Medical Device Epidemiological Network Initiative (MDEpiNet).Am Heart J. 2016; 171:64.e1–72.e1. doi: 10.1016/j.ahj.2015.07.029CrossrefGoogle Scholar6. Selzman KA. EP PASSION: A new paradigm for lead surveillance.Oral presentation at: Heart Rhythm Society Scientific Sessions; May 10, 2019; San Francisco, CA.Google Scholar7. Abbott. Product Performance Report.https://www.cardiovascular.abbott/content/dam/bss/divisionalsites/cv/pdf/reports/abbott-second-edition-ppr-2019.pdf. Accessed March 20, 2020.Google Scholar8. Boston Scientific. Rhythm Management Product Performance Report.https://www.bostonscientific.com/content/dam/bostonscientific/quality/ppr/2018/Q2/Product%20Performance%20Report%20Q2%202018%20Rev%20A.pdf. Accessed March 20, 2020.Google Scholar9. Medtronic. CRHF Product Performance eSource.https://wwwp.medtronic.com/productperformance/. Accessed March 20, 2020.Google Scholar10. Biotronik. Product Performance Report.https://biotronik.cdn.mediamid.com/cdn_bio_doc/bio32145/85779/bio32145.pdf. Accessed March 20, 2020.Google Scholar11. Jensen T, Chin J, Ashby L, Dolan D. Decision Memo for Implantable Automatic Defibrillators (CAG-00157R4). Centers for Medicare and Medicaid Services; 2018. https://www.cms.gov/medicare-coverage-database/details/nca-decision-memo.aspx?NCAId=288Google Scholar Previous Back to top Next FiguresReferencesRelatedDetailsCited By Gasperetti A, Schiavone M, Ziacchi M, Vogler J, Breitenstein A, Laredo M, Palmisano P, Ricciardi D, Mitacchione G, Compagnucci P, Bisignani A, Angeletti A, Casella M, Picarelli F, Fink T, Kaiser L, Hakmi S, Calò L, Pignalberi C, Santini L, Lavalle C, Pisanò E, Olivotto I, Tondo C, Curnis A, Dello Russo A, Badenco N, Steffel J, Love C, Tilz R, Forleo G and Biffi M (2021) Long-term complications in patients implanted with subcutaneous implantable cardioverter-defibrillators: Real-world data from the extended ELISIR experience, Heart Rhythm, 10.1016/j.hrthm.2021.07.008, 18:12, (2050-2058), Online publication date: 1-Dec-2021. Joglar J (2021) Electrical abnormalities with St. Jude Medical/Abbott pacing leads: Let's not call it lead failure yet, Heart Rhythm, 10.1016/j.hrthm.2021.09.002, 18:12, (2070-2071), Online publication date: 1-Dec-2021. Khatiwala R, Mullins E, Fan D, Srivatsa U, Dhruva S and Oesterle A (2021) Electrical abnormalities with St. Jude/Abbott pacing leads: A systematic review and meta-analysis, Heart Rhythm, 10.1016/j.hrthm.2021.08.026, 18:12, (2061-2069), Online publication date: 1-Dec-2021. Dhruva S, Zeitler E, Caños D, Odobasic H, Major J, Wang L, Mullane S, Nabutovsky Y, Smith V, Zhang L and Selzman K (2022) Evaluation of cardiovascular implantable electronic device leads post implant: ElectroPhysiology Predictable And SuStainable Implementation Of National Registries (EP PASSION), Journal of Interventional Cardiac Electrophysiology, 10.1007/s10840-022-01189-x Related articlesActive Surveillance of the Implantable Cardioverter-Defibrillator Registry for Defibrillator Lead FailuresFrederic S. Resnic, et al. Circulation: Cardiovascular Quality and Outcomes. 2020;13 April 2020Vol 13, Issue 4 Advertisement Article InformationMetrics © 2020 American Heart Association, Inc.https://doi.org/10.1161/CIRCOUTCOMES.120.006649PMID: 32283972 Originally publishedApril 14, 2020 Keywordsregistriesproduct surveillance, postmarketingdefibrillator, implantabledevice approvalEditorialsalgorithmsPDF download Advertisement SubjectsEthics and PolicyQuality and Outcomes
No takes yet. Share an insight, caveat, or question.
Zeitler et al. (2020) conducted an editorial in Implantable cardioverter-defibrillator (ICD) implantation (n=370,000). Specific ICD lead models vs. Propensity-matched comparator group receiving all other lead models was evaluated on Lead failure (using lead replacement as a surrogate). Active surveillance of over 370,000 ICD leads using the NCDR registry found no difference in lead survival rates between matched groups, with failure-free survival ranging from 97.7% to 98.9%.
Synapse has enriched 10 related papers. Consider them for comparative context: