Key result
Implantation of a continuous-flow left ventricular assist device in a patient with a subcutaneous ICD resulted in 31 inappropriate shocks due to electromagnetic interference and oversensing.
Population
1 42-year-old man with nonischemic cardiomyopathy, severely reduced left ventricular function, and advanced…
Design
Case_report
Follow-up
6 months
Authors
Loading...
May prompt S-ICD deactivation pre-LVAD; single case leaves open optimal perioperative protocols.
Case Report (n=1)
No
LVAD implantation in a patient with an S-ICD can cause severe electromagnetic interference and inappropriate shocks, highlighting the need to deactivate antitachycardia therapy before surgery and carefully evaluate sensing vectors postoperatively.
Pfeffer et al. (2016) conducted a case report in Nonischemic cardiomyopathy and advanced heart failure (n=1). Left ventricular assist device (LVAD) implantation was evaluated on Inappropriate S-ICD shocks. Implantation of a continuous-flow left ventricular assist device in a patient with a subcutaneous ICD resulted in 31 inappropriate shocks due to electromagnetic interference and oversensing.