Answers are grounded in this paper and cite their sources.
We developed a nonisotopic technique, Hepagene, for measuring hepatitis B virus (HBV) DNA in human serum by using a sulfonated probe that is detected by a sandwich immunoenzymatic reaction. The detection limit, determined by serum dilution tests, was 2.5 ng/L. The precision of the Hepagene test was demonstrated by the accurate reproducibility observed for low (3 ng/L) and medium (38 ng/L) concentrations of HBV DNA assayed in 24 different series. Specificity was established by assaying HBV DNA in sera from 98 patients by the Hepagene technique or by a solution hybridization assay with an 125I-labeled probe. Results by both techniques agreed for 94 sera (96%), with 68 being concordant for HBV DNA negativity and 26 for positivity. HBV DNA titers assayed by both methods also agreed. Hepagene represents the first nonisotopic HBV DNA assay involving a sulfonated probe and with performance characteristics equivalent to those of classical radioactive hybridization techniques.
No takes yet. Share an insight, caveat, or question.
AKAR et al. (1992) studied this question.