analyses of the primers and probes and wet testing. The SAMBA II SARS-CoV-2 test was evaluated for performance characteristics. Clinical performance was evaluated in residual combined throat/nose swabs and compared to that of the Public Health England real-time PCR assay targeting the RdRp gene. The SAMBA II SARS-CoV-2 test has an analytical sensitivity of 250 copies/ml for detecting two regions of the genome (open reading frame 1ab [ORF1ab] and nucleocapsid protein [N]). The clinical performance was evaluated in 172 residual combined nose/throat swabs provided by the Clinical Microbiology and Public Health Laboratory, Addenbrooke's Hospital, Cambridge (CMPHL), which showed an estimated positive percent agreement of 98.9% (95% confidence interval [CI], 93.83 to 99.97) and negative percent agreement of 96.4% (95% CI, 89.92 to 99.26) compared to testing by the CMPHL. The data show that the SAMBA II SARS-CoV-2 test performs equivalently to the centralized testing methods, but with a shorter turnaround time of 86 to 101 min. Point-of-care tests such as SAMBA should enable rapid patient management and effective implementation of infection control measures.
No takes yet. Share an insight, caveat, or question.
Assennato et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: