Key result
TTE identifies a vegetation-like structure on a mechanical aortic valve as degassing microbubbles, not endocarditis.
Why the study?
Mechanical heart valve vegetations can be misinterpreted on echocardiography due to the presence of degassing microbubbles, complicating diagnosis of infective endocarditis.
Case Report (n=1)
Degassing microbubbles on mechanical heart valves can mimic vegetations on echocardiography and should be considered in the differential diagnosis of suspected infective endocarditis.
Answers are grounded in this paper and cite their sources.
Consider degassing microbubbles in suspected mechanical valve endocarditis differentials; this case report leaves open prevalence and diagnostic criteria.
A 17-year-old male with a history of newly implanted mechanical valve at the aortic position, presented with fever, rigors, and painful cutaneous abscesses on his lower extremities and was suspected for infective endocarditis. Transthoracic echocardiography (TTE) showed a vegetation-like structure following the movement of the mechanical heart valve (MHV), which eventually proved to be a product of degassing microbubbles (MB).
No takes yet. Share an insight, caveat, or question.
Lerche et al. (2013) conducted a case report in Suspected infective endocarditis (n=1). Degassing microbubbles was evaluated on Echocardiographic findings. A vegetation-like structure observed on transthoracic echocardiography in a patient with a mechanical aortic valve was identified as degassing microbubbles rather than infective endocarditis.