Key result
Newer Navion stent grafts show no significant difference in aortic PWV versus older Captivia grafts.
Why the study?
Little is known about cardiovascular changes after TEVAR or the impact of different stent graft generations and design changes on aortic stiffness.
Does a newer generation thoracic aortic stent graft (Navion) reduce stent graft-induced aortic stiffening compared to an older generation (Captivia) in an ex vivo porcine model?
Does a newer generation thoracic aortic stent graft (Navion) reduce stent graft-induced aortic stiffening compared to an older generation (Captivia) in an ex vivo porcine model?
Both generations of thoracic aortic stent grafts significantly increased aortic stiffness (measured by PWV) in an ex vivo model, with no significant difference between the older and newer device designs.
Answers are grounded in this paper and cite their sources.
Newer and older thoracic stent grafts increase aortic stiffness similarly in this ex vivo model; hypothesis-generating and should not yet change practice.
Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the Valiant thoracic aortic stent graft. This was an ex vivo porcine investigation using an experimental mock circulatory loop. Thoracic aortas of young healthy pigs were harvested and connected to the mock circulatory loop. At a 60 bpm heart rate and stable mean arterial pressure, baseline aortic characteristics were obtained. Pulse wave velocity (PWV) was calculated before and after stent graft deployment. Paired and independent sample t tests or their non-parametric alternatives were performed to test for differences where appropriate. Twenty porcine thoracic aortas were divided into two equal subgroups, in which a Valiant Captivia or a Valiant Navion stent graft was deployed. Both stent grafts were similar in diameter and length. Baseline aortic characteristics did not differ between the subgroups. Mean arterial pressure values did not change after either stent graft, while pulse pressures increased statistically significantly after Captivia (mean 44 ± 10 mmHg to 51 ± 13 mmHg, p = .002) but not after Navion. Mean baseline PWV increased after both Captivia (4.4 ± 0.6 m/s to 4.8 ± 0.7 m/s, p = .007) and Navion (4.6 ± 0.7 m/s to 4.9 ± 0.7 m/s, p = .002). There was no statistically significant difference in the mean percentage increase in PWV for either subgroup (8 ± 4% vs. 6 ± 4%, p = .25). These experimental findings showed no statistically significant difference in the percentage increase of aortic PWV after either stent graft generation and confirm that TEVAR increases aortic PWV. As a surrogate for aortic stiffness, this calls for further improvements in future thoracic aortic stent graft designs regarding device compliance.
No takes yet. Share an insight, caveat, or question.
Mandigers et al. (2023) studied Healthy porcine aortas (ex vivo model for TEVAR) (n=20). Valiant Navion stent graft vs. Valiant Captivia stent graft was evaluated on Mean percentage increase in pulse wave velocity (PWV) (p=0.25). Deployment of the newer generation Navion stent graft resulted in a 6% mean increase in aortic pulse wave velocity compared to an 8% increase with the older Captivia graft, a difference that was not significant.