Key result
Pacemaker-induced RNRVAS causes loss of optimal AV delay, inappropriate ventricular pacing, and inaccurate AHRE diagnosis.
Why the study?
Pacemaker-induced arrhythmias such as RNRVAS must be distinguished from atrial fibrillation due to their clinical impact and diagnostic challenges.
This review highlights the importance of distinguishing pacemaker-induced arrhythmias like RNRVAS from atrial fibrillation to prevent inappropriate device behavior and inaccurate diagnosis.
Answers are grounded in this paper and cite their sources.
RNRVAS must be distinguished from AF in device-detected AHRE; leaves open standardized diagnostic criteria for pacemaker programming.
Repetitive non‐reentrant ventriculo‐atrial synchrony (RNRVAS) is a pacemaker‐induced arrhythmia that must be distinguished from atrial fibrillation (AF). Pacemaker‐induced arrhythmias are commonly detected as atrial high rate episodes (AHRE) by implanted cardiac devices. Two main types of atrial oversensing are recognized: far‐field R‐wave (FFRW) oversensing and pacemaker‐induced arrhythmias, which include pacemaker‐mediated tachycardia and RNRVAS. The presence of ventriculo‐atrial conduction is required for both types of pacemaker‐induced arrhythmias. The incidence of RNRVAS can increase with the use of various device settings and functions, such as long atrioventricular (AV) interval programming, the rate‐adaptive mode, and the atrial overdrive pacing algorithm. The negative aspects of pacemaker‐induced arrhythmias, especially RNRVAS, include (1) loss of optimal AV delay, (2) inappropriate increase in ventricular pacing, (3) induction of atrial arrhythmias, and (4) inaccurate diagnosis of AHRE. We review the incidence of arrhythmias, electrophysiological mechanisms, and the clinical diagnosis of RNRVAS identified by using dual‐chamber implantable cardiac devices.
No takes yet. Share an insight, caveat, or question.
Kohno et al. (2014) conducted a review in Pacemaker-induced arrhythmias (RNRVAS). Dual-chamber implantable cardiac devices was evaluated. Repetitive non-reentrant ventriculo-atrial synchrony (RNRVAS) is a pacemaker-induced arrhythmia that can cause loss of optimal AV delay, inappropriate ventricular pacing, and inaccurate AHRE diagnosis.