Answers are grounded in this paper and cite their sources.
A direct contact applicator, specifically designed for microwave diathermy at the Industrial, Scientific Medical (ISM) frequency of 2.45 GHz was evaluated by studying near-field patterns in free space, thermographic heating patterns in phantoms of simulated fat and muscle tissue, and associated leakage radiation. The main features are a circular waveguide with a short conical flare horn output section surrounded by an annular choke and two sets of dual posts to generate far-field circular polarization. The significant near field components of the therapeutic beam are in a transverse plane, parallel to the aperture. Heating patterns on the exposed surface of muscle phantoms and inside fat-muscle phantoms are spatially similar and relatively uniform. The leakage level is 0.8 mW/cm/sup 2/ per 100 W of forward power for direct contact and 4 mW/cm/sup 2/ per 100 W of forward power for a 1-cm air gap between aperture and planar phantoms. The uncertainty of these leakage measurements is /spl plusmn/2 dB. This investigation demonstrates the technical feasibility of a safe and effective direct contact microwave diathermy applicator operating at 2.45 GHz. The applicator is a viable candidate for hyperthermia appllications.
No takes yet. Share an insight, caveat, or question.
Kantor et al. (1978) studied this question.