Key result
Zener mathematical model accurately predicts up to ~5% pressure wave attenuation in viscoelastic prostheses.
Why the study?
The effect of viscoelastic polymeric materials used in vascular prostheses on blood pulse propagation remains to be fully understood.
Mathematical modeling of viscoelastic materials demonstrates how vascular prostheses can be tailored to attenuate and damp arterial blood pulses.
Modeling suggests viscoelastic prostheses may damp pulses; leaves open clinical translation and outcomes.
The present study examines the possibility of attenuating blood pulses by means of introducing prosthetic viscoelastic materials able to absorb energy and damp such pulses. Vascular prostheses made of polymeric materials modify the mechanical properties of blood vessels. The effect of these materials on the blood pulse propagation remains to be fully understood. Several materials for medical applications, such as medical polydimethylsiloxane or polytetrafluoroethylene, show viscoelastic behavior, modifying the original vessel stiffness and affecting the propagation of blood pulses. This study focuses on the propagation of pressure waves along a pipe with viscoelastic materials using the Maxwell and the Zener models. An expression of exponential decay has been obtained for the Maxwell material model and also for low viscous coefficient values in the Zener model. For relatively high values of the viscous term in the Zener model, the steepest part of the pulse can be damped quickly, leaving a smooth, slowly decaying wave. These mathematical models are critical to tailor those materials used in cardiovascular implants to the mechanical environment they are confronted with to repair or improve blood vessel function.
No takes yet. Share an insight, caveat, or question.
Menacho et al. (2017) studied Arterial pulse attenuation. Viscoelastic material prosthesis (PDMS) was evaluated on Pressure wave attenuation. The Zener mathematical model accurately predicted pressure wave attenuation in a viscoelastic polydimethylsiloxane prosthesis, demonstrating up to 5% attenuation in a 200-mm-long cylinder.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: