Key result
MoyoAssist Extra-VAD provides safe cardiac support in sheep over 30 days without device-related adverse events.
Why the study?
Extracorporeal ventricular assist devices have been used for cardiogenic shock, and this study aimed to evaluate the 30-day in-vivo performance and safety of the newly developed MoyoAssist Extra-VAD.
Does the MoyoAssist Extra-VAD provide safe and effective cardiac support in a healthy ovine model over 30 days?
Does the MoyoAssist Extra-VAD provide safe and effective cardiac support in a healthy ovine model over 30 days?
The MoyoAssist Extra-VAD demonstrated safety and effective hemodynamic support in a 30-day ovine model, supporting its potential for clinical translation in cardiogenic shock.
OBJECTIVE: Cardiogenic shock (CS) often occurs in patients suffering from rapidly progressing end-stage heart failure or acute myocardial infarction. Mechanical circulatory support may be used for patients who do not respond to medication or revascularization to stabilize hemodynamics. Extracorporeal ventricular assist device (Extra-VAD) has been reported to be successful for patients with cardiogenic shock. This study aimed to evaluate the 30-day in-vivo performance and safety of a newly developed Extra-VAD with maglev centrifugal pump technology, MoyoAssist®. METHOD: The study was conducted with 6 healthy ovine models, weighing 43.2 ~ 59.6 kg. Cannulation was performed with a 34 Fr venous cannula surgically connected to the left arterial appendage and a 24 Fr arterial cannula inserted into descending aorta. The pump flow rate was maintained at 2 ~ 3 L/min to provide sufficient cardiac support without suction. Activated clotting time was maintained within the range of 150 ~ 250 s. RESULTS: No device-related adverse events occurred throughout the study. Plasma-free hemoglobin results were within the acceptable range of ventricular assist device therapy (<40 mg/dl). MGS01 had an anticoagulation management related bleeding event and was terminated on day 29. All other sheep's biochemical test results were stable. The autopsy showed no embolism or thrombus formation and no end-organ damage. CONCLUSION: This study demonstrated that the MoyoAssist® Extra-VAD is able to provide cardiac support effectively and safely and may provide a new alternative choice for patients with CS in China.
No takes yet. Share an insight, caveat, or question.
Li et al. (2022) studied Healthy ovine models (for cardiogenic shock) (n=6). MoyoAssist® Extra-VAD was evaluated on 30-day in-vivo performance and safety. The MoyoAssist Extra-VAD provided effective and safe cardiac support in 6 healthy sheep over 30 days, with no device-related adverse events and acceptable plasma-free hemoglobin levels (<40 mg/dl).