This feasibility study evaluates safety and hemodynamic performance of CorVad 6.0 device in sheep, implying potential for future use in heart failure treatment.
This study aimed to evaluate the feasibility and safety of a novel percutaneous left ventricular assist device CorVad 6.0, for mechanical circulatory support in an ovine model, focusing on device performance, hemocompatibility, and end-organ effects. The CorVad 6.0, which is a microaxial flow pump incorporating an integrated axial-flux motor, was implanted in six healthy sheep via descending aortic access. Animals were supported for 4 weeks, with pump speeds titrated to maintain flows of 1.5–5.0 L/min. All six animals survived the 4 week study period. The CorVad 6.0 was successfully implanted in all subjects with no device-related complications, demonstrating stable operation and a predictable flow response to speed changes. Key hematological and biochemical parameters, including plasma-free hemoglobin, remained within acceptable ranges throughout the study, showing no evidence of significant hemolysis or end-organ dysfunction. Macroscopic and histological analyses of the heart, liver, kidneys, and brain revealed no device-related pathological abnormalities. The CorVad 6.0 demonstrates stable hemodynamic performance and a favorable biocompatibility during a 4 week implantation period. Further study investigating chronic heart failure modes is needed.