Key result
The Penn State Infant Ventricular Assist Device achieved the 4-6 weeks survival goal in 5 of 7 juvenile sheep, with a mean support duration of 26.1 days and minimal device thrombi.
Why the study?
Does the Penn State Infant VAD provide safe circulatory support in a juvenile ovine model?
Does the Penn State Infant VAD provide safe circulatory support in a juvenile ovine model?
The Penn State Infant VAD demonstrated feasibility and safety in a juvenile ovine model for up to 41 days, with TEG-guided anticoagulation reducing heparin requirements without increasing thromboembolism.
Supports preclinical validation of infant VADs; leaves open translation to human neonates.
The Penn State Infant Ventricular Assist Device (VAD) is a 12-14 ml stroke volume pneumatically actuated pump, with custom Björk-Shiley monostrut valves, developed under the National Heart, Lung, and Blood Institute Pediatric Circulatory Support program. In this report, we describe the seven most recent chronic animal studies of the Infant VAD in the juvenile ovine model, with a mean body weight of 23.5 ± 4.1 kg. The goal of 4-6 weeks survival was achieved in five of seven studies, with support duration ranging from 5 to 41 days; mean 26.1 days. Anticoagulation was accomplished using unfractionated heparin, and study animals were divided into two protocol groups: the first based on a target activated partial thromboplastin time of 1.5-2 times normal, and a second group using a target thromboelastography R-time of two times normal. The second group required significantly less heparin, which was verified by barely detectable heparin activity (anti-Xa). In both groups, there was no evidence of thromboembolism except in one animal with a chronic infection and fever. Device thrombi were minimal and were further reduced by introduction of the custom valve. These results are consistent with results of adult VAD testing in animals and are encouraging given the extremely low levels of anticoagulation in the second group.
No takes yet. Share an insight, caveat, or question.
Weiss et al. (2011) studied Pediatric circulatory support (animal model) (n=7). Penn State Infant Ventricular Assist Device was evaluated on 4-6 weeks survival. The Penn State Infant Ventricular Assist Device achieved the 4-6 weeks survival goal in 5 of 7 juvenile sheep, with a mean support duration of 26.1 days and minimal device thrombi.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: