Key result
Operating the 12 cc Penn State pulsatile pediatric VAD with a 50 ms end-diastolic delay provided superior wall shear profiles compared to 10 ms and 100 ms delays due to a longer sustained inlet jet.
Why the study?
Does altering the end-diastolic delay time improve flow fields and wall shear rates in the 12 cc Penn State pulsatile pediatric VAD?
Does altering the end-diastolic delay time improve flow fields and wall shear rates in the 12 cc Penn State pulsatile pediatric VAD?
Answers are grounded in this paper and cite their sources.
In a pediatric pulsatile VAD model, an end-diastolic delay of 50 ms optimizes fluid dynamics and wall shear rates, potentially reducing thrombosis risk during weaning.
The success of adult ventricular assist devices (VADs), coupled with the high transplant waiting list mortality of infants (40%) has prompted Penn State to develop a pediatric version of the clinically successful adult device. Although the primary use of this device will be bridge-to-transplant, there has been sufficient clinical data to demonstrate the efficacy of VADs in a bridge-to-recovery setting. However, removing the patient from the device, a process known as weaning, demands operation of the device at a lower beat rate and concomitant increased risk for thromboembolism. Previous studies have shown that the interrelated flow characteristics necessary for the prevention of thrombosis in a pulsatile VAD are a strong inlet jet, a late diastolic recirculating flow, and a wall shear rate greater than 500/s. In an effort to develop a strong inlet jet and rotational flow pattern at a lower beat and flow rate, we have compressed diastole by altering the end-diastolic delay time (EDD). Particle image velocimetry was used to compare the flow fields and wall shear rates in the chamber of the 12 cc Penn State pulsatile pediatric VAD operated at 50 beats per minute using EDDs of 10, 50, and 100 ms. Although we expected the 100 ms EDD to have the best wall shear profiles, we found that the 50 ms EDD condition was superior to both the 10 and 100 EDD conditions, due to a longer sustained inlet jet.
No takes yet. Share an insight, caveat, or question.
Cooper et al. (2010) studied this question. End-diastolic delay (EDD) of 50 ms vs. EDDs of 10 and 100 ms was evaluated on Flow fields and wall shear rates. Operating the 12 cc Penn State pulsatile pediatric VAD with a 50 ms end-diastolic delay provided superior wall shear profiles compared to 10 ms and 100 ms delays due to a longer sustained inlet jet.