Answers are grounded in this paper and cite their sources.
Demonstrates passive circulation balancing in continuous-flow total artificial hearts via hydromechanical pressure regulation, highlighting sensor-free control.
Control of mechanical circulatory support pump output typically requires that pressure-regulating functions be accomplished by active control of the speed or geometry of the device, with feedback from pressure or flow sensors. This article presents a different design approach, with a pressure-regulating device as the core design feature, allowing the essential control function of regulating pressure to be directly programmed into the hydromechanical design. We show the step-by-step transformation of a pressure-regulating device into a continuous-flow total artificial heart that passively balances left and right circulations without the need for pressure and flow sensors. In addition, we discuss a ventricular assist device that prevents backflow in the event of power interruption and also dynamically interacts with residual ventricle function to preserve pulsatility.
No takes yet. Share an insight, caveat, or question.
Horvath et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: