Key result
Mini-electrode ablation cuts ablation time ~64% and reduces attempts vs conventional catheters in PSVT.
Why the study?
Catheters equipped with mini-electrodes provide high-resolution mapping, but their mapping performance and efficacy during radiofrequency ablation of PSVTs needed evaluation.
Does a mini-electrode catheter reduce ablation attempts and time in patients undergoing radiofrequency ablation for paroxysmal supraventricular tachycardia?
RCT (n=136)
randomized
Does a mini-electrode catheter reduce ablation attempts and time in patients undergoing radiofrequency ablation for paroxysmal supraventricular tachycardia?
Absolute Event Rate: 2% vs 3%
p-value: p=0.032
Mini-electrode catheters improve mapping resolution and significantly reduce the number of ablation attempts and time required to achieve procedural endpoints in PSVT ablation.
ME catheters may enhance mapping in PSVT ablation; leaves open whether they improve procedural success or durability versus conventional catheters.
Introduction: Novel ablation catheters equipped with mini-electrodes (ME) offer high resolution mapping for target tissue. This study aimed to evaluate the mapping performance and efficacy of ME catheters in radiofrequency ablation of paroxysmal supraventricular tachycardias (PSVTs). Methods: We prospectively enrolled 136 patients who were undergoing catheter ablation of PSVT including 76 patients with atrioventricular nodal reentrant tachycardia (AVNRT) and 60 patients with atrioventricular reentrant tachycardia (AVRT) or Wolff-Parkinson-White (WPW) syndrome. Patients were randomized to the ME group (ablation using a 4.5mm tip ME catheter) or the control group (ablation using a conventional 4mm tip catheter). The number of ablation attempt and cumulative ablation time to ablation endpoints, which was defined as an emergence of junctional rhythm in AVNRT or accessory pathway (AP) block in AVRT/WPW syndrome were compared. Results: During ablation procedures, discrete SP or AP electrograms were found in 27 (39.7%) patients in the ME group and 13 (19.1%) patients in the control group. The primary study outcomes were significantly lower in the ME group (ablation attempt number: 2.0 [1–4] vs. 3.0 [2–7] in the ME and control group, p=0.032; ablation time: 23.5 [5.0–111.5] vs. 64.5 [16.0–185.0] seconds, p=0.013). According to the PSVT diagnosis, ablation time to junctional rhythm was significantly shorter in the ME group in AVNRT. In AVRT/WPW syndrome, both ablation attempt number and ablation time to AP block showed trends favoring the ME group. Conclusion: The novel ME catheter was advantageous for identifying pathway potentials and reducing initial ablation attempt number and time in PSVT ablation. (ClinicalTrials.gov number, NCT04215640)
No takes yet. Share an insight, caveat, or question.
Choi et al. (2022) conducted an RCT in Paroxysmal supraventricular tachycardia (PSVT) (n=136). Mini-electrode (ME) catheter vs. Conventional 4mm tip catheter was evaluated on Number of ablation attempts to ablation endpoints (p=0.032). Ablation using a mini-electrode catheter significantly reduced the median number of ablation attempts (2.0 vs 3.0, p=0.032) and ablation time (23.5 vs 64.5 seconds, p=0.013) in patients with PSVT.