Answers are grounded in this paper and cite their sources.
Optical fibre dosemeter systems based on radioluminescence and optically stimulated luminescence (OSL) from carbon-doped aluminium oxide (Al(2)O(3):C) crystals were developed for in vivo real-time dose rate and absorbed dose measurements in radiotherapy and mammography. A technique was also developed for making ultra-small dosemeter probes that can easily be placed inside patients in radiation treatment. These probes have shown excellent properties in both head and neck intensity-modulated radiation therapy treatment and in mammography. The dose-response of the OSL signal for the new optical fibre dosemeter system showed a repeatability within 0.15% at a dose level of 60 mGy when integrated over 100 s. The temperature dependence in the range 0-45 degrees C showed a reproducibility within 1.3% when the OSL signal was integrated over 100 s.
No takes yet. Share an insight, caveat, or question.
Marckmann et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: