Key result
Sphygmetrics SR-2 automatic BP recorder correlates highly with mercury sphygmomanometry for systolic and diastolic measurements.
Why the study?
The Sphygmetrics SR-2 Automatic Blood Pressure Recorder may reduce observer bias in epidemiologic blood pressure studies but requires comparison to the mercury sphygmomanometer.
Does the Sphygmetrics SR-2 Automatic Blood Pressure Recorder provide comparable blood pressure measurements to a mercury sphygmomanometer?
Cross-Sectional (n=71)
Does the Sphygmetrics SR-2 Automatic Blood Pressure Recorder provide comparable blood pressure measurements to a mercury sphygmomanometer?
Answers are grounded in this paper and cite their sources.
The Sphygmetrics SR-2 Automatic Blood Pressure Recorder is comparable to auscultation for estimating systolic BP, reducing observer bias, but may be less reliable for diastolic BP in thin individuals.
The Sphygmetrics SR-2 Automatic Blood Pressure Recorder uses an infrasonic technique for detecting artery wall motion to estimate systemic arterial pressure and produces a permanent record of the results. It therefore is potentially useful in reducing observer bias in epidemiologic studies of blood pressure (BP). Two blood pressures were recorded in 21 men and 50 women using the SR-2 simultaneously with two auscultators using a biaural stethoscope and mercury syhgmomanometer. The SR-2 measured slightly higher systolic nd slightly lower diastolic pressures on average, but the differences were not significant. The two auscultators were highly correlated with one another (r = 0.99 systolic/0.97 diastolic) and with the SR-2 (0.93 and 0.92 for systolic, 0.84 and 0.85 for diastolic). The correlations were unrelated to sex, age, or antihypertensive medication status, but the correlations between th SR-2 and either auscultator for diastolic BP were quite low (0.36, 0.55) in subjects whose relative weight was below 1.0. Interpretation of the SR-2 disc was very reliable, with inter- and intra-reader correlations being 0.99 for systolic and 0.84 to 0.94 for diastolic. The SR-2 was found to be comparable to auscultation in estimating systolic BP in a heterogeneous population which has the advantages of reducing observer bias and producing a permanent record. Its use in estimating diastolic pressure in thin individuals and children needs further evaluation.
No takes yet. Share an insight, caveat, or question.
Fortmann et al. (1981) conducted a cross-sectional in Blood pressure measurement (n=71). Sphygmetrics SR-2 Automatic Blood Pressure Recorder vs. Mercury sphygmomanometer (auscultation) was evaluated on Correlation of systolic and diastolic blood pressure measurements (r = 0.93 and 0.92 for systolic, 0.84 and 0.85 for diastolic). The Sphygmetrics SR-2 Automatic Blood Pressure Recorder showed high correlation with mercury sphygmomanometry for systolic and diastolic blood pressure measurements.
Synapse has enriched 10 related papers. Consider them for comparative context: