Key result
Automated blood pressure measurements taken while the patient rested alone (142/80 mmHg) were significantly lower than manual mercury sphygmomanometer readings (163/86 mmHg, P<0.001).
Why the study?
Does automated blood pressure measurement reduce the white-coat effect compared to a conventional mercury sphygmomanometer in patients with hypertension?
Population
50 consecutive patients referred to a specialist for management of hypertension.
Comparison
Automated blood pressure measurement using the… vs Conventional mercury sphygmomanometer.
Design
Cross-sectional, random order
Authors
Loading...
Automated BP may reduce white-coat effect; leaves open need for outcome trials before practice change.
RCT (n=50)
Random order
Does automated blood pressure measurement reduce the white-coat effect compared to a conventional mercury sphygmomanometer in patients with hypertension?
Absolute Event Rate: 142% vs 163%
p-value: p=<0.001
Automated blood pressure measurement taken while the patient rests alone significantly reduces the white-coat effect compared to manual mercury sphygmomanometer readings.
Martin G. Myers (2006) conducted an RCT in Hypertension (n=50). Automated blood pressure recorder (BpTRU) vs. Conventional mercury sphygmomanometer was evaluated on Mean systolic blood pressure (mmHg) (p=<0.001). Automated blood pressure measurements taken while the patient rested alone (142/80 mmHg) were significantly lower than manual mercury sphygmomanometer readings (163/86 mmHg, P<0.001).