We demonstrate dual-comb near-infrared spectroscopy (DC-fNIRS) in reflectance geometry using an electro-optic architecture with optical injection locking that requires no ultrafast detectors, no photon-counting hardware, and no sequential time scanning. Two mutually coherent optical frequency combs are generated from a single continuous-wave laser at 852 nm and interrogate the medium in parallel, down-converting its complex frequency response to the radio-frequency domain, where standard acquisition electronics can be used. Applied to a biomimetic phantom replicating the optical properties of human brain tissue, the system reconstructs the full diffuse time-of-flight distribution (DTOF) with 250 ps path-length resolution and a DTOF update rate of ∼10,000 complete reconstructions per second ( T update = 100~ μ s), and retrieves absolute absorption and reduced scattering coefficients ( μ a , μ s ′). A perturbation transfer function isolates the contribution of a localized subsurface inclusion—a carbon fiber rod of 1.5 mm diameter—at 1.5 cm depth, demonstrating path-length-resolved sensitivity to clinically relevant inhomogeneities without prior calibration. The electro-optic architecture allows the number of comb lines, spectral span, and update rate to be adjusted in software, providing a scalable route toward compact, high-density neuroimaging systems for functional brain monitoring.