Key result
Noncontact mapping yields ~24 ms activation timing errors during ventricular fibrillation, exceeding sinus rhythm errors.
Why the study?
The utility of noncontact mapping in acute ischemic ventricular fibrillation has not been directly compared with simultaneous multisite contact needle recordings.
Does noncontact mapping accurately reflect endocardial electrical activation compared to contact recordings in acute ischemic ventricular fibrillation in a sheep model?
Does noncontact mapping accurately reflect endocardial electrical activation compared to contact recordings in acute ischemic ventricular fibrillation in a sheep model?
Noncontact mapping exhibits significant errors in activation timing and sequence during ventricular fibrillation compared to sinus rhythm, suggesting limitations in its utility for mapping chaotic rhythms like VF.
Noncontact mapping timing errors increase in VF; leaves open its reliability for arrhythmia mapping and requires contact-validated refinement.
AIM: The study aim was to determine the utility of noncontact mapping in acute ischemic ventricular fibrillation, by direct comparison with simultaneously acquired multisite contact needle recordings. MATERIALS AND METHODS: Noncontact mapping has emerged as a promising tool to make percutaneous high-density intracardiac electrical recordings in clinical and research settings. Previous large-animal mapping studies of ischemic VF have used electrodes in contact with the fibrillating myocardium. A previous report of noncontact mapping in VF used a single intracardiac catheter to validate noncontact recordings. RESULTS: We studied the spontaneous onset of VF in 8 sheep, after acute occlusion of the left anterior descending coronary artery. Simultaneous unipolar recordings of endocardial electrical activation were made with the most endocardial electrode of 32 transmural plunge needles (contact electrograms; CE), and an Ensite noncontact catheter in the left ventricle (noncontact virtual electrograms; NCVE). The lag-shifted morphology cross-correlation between individual NCVE and CE pairs during VF was calculated. DISCUSSION: The mean lag-shifted cross-correlation coefficient during manually selected sinus rhythm beats was 0.89 ± 0.006, and 0.80 ± 0.009 in VF (P < 0.05). The mean absolute time shift was 7.3 ± 0.5 ms in sinus rhythm, and 23.8 ± 0.9 ms in VF (P < 0.001). Fast Fourier transform demonstrated well-correlated dominant frequencies between CE and NCVE. VF myocardial activation maps of NCVE showed organized wavefronts, but dyssynchronous activation in CE. CONCLUSION: These data demonstrate errors in activation timing and sequence with noncontact mapping more prominent during VF than sinus rhythm.
No takes yet. Share an insight, caveat, or question.
Ganesan et al. (2010) studied acute ischemic ventricular fibrillation (n=8). Noncontact mapping (Ensite noncontact catheter) vs. Multisite contact needle recordings was evaluated on Lag-shifted morphology cross-correlation and absolute time shift between noncontact and contact electrograms (p=<0.001). Noncontact mapping demonstrated greater errors in activation timing compared to contact mapping during ventricular fibrillation (time shift 23.8 ms) than in sinus rhythm (7.3 ms, P<0.001).