Key result
Varying Doppler insonation angle alters ICA peak systolic velocity, with greatest variance at 60°.
Why the study?
Doppler velocity measurements rely on insonation angles of 60 degrees or less to minimize error, prompting an assessment of peak systolic velocity variance in internal carotid arteries across angles of 45, 50, 55, and 60 degrees with different beam steering.
Does the angle of insonation affect Doppler-detected peak systolic velocity measurements in the internal carotid arteries of asymptomatic volunteers?
Cross-Sectional (n=22)
Does the angle of insonation affect Doppler-detected peak systolic velocity measurements in the internal carotid arteries of asymptomatic volunteers?
p-value: p=<.01
Doppler-detected peak systolic velocity measurements in the internal carotid artery vary significantly depending on the angle of insonation, highlighting the need for consistent ultrasound techniques.
OBJECTIVES: Doppler velocity measurements are fundamental diagnostic criteria for vascular ultrasound examinations. Insonation angles are kept to 60° or less to minimize error. The purpose of this study was to assess variance of Doppler-detected peak systolic velocity (PSV) measurements in the internal carotid arteries at different angles (45°, 50°, 55°, and 60°) with different beam steering. METHODS: The PSV was recorded from the right and left internal carotid arteries in 22 asymptomatic volunteers with straight vessels (total of 44 vessels). A standardized approach was used for recording velocities with the Doppler cursor center steered and steered 15° from right to left. An analysis of variance was performed. RESULTS: The PSV varied significantly with the 4 different angles of insonation (P < .01). The maximum variation between 45° and 60° angles within a single vessel was 29 cm/s. The average variation over the 4 angles was 14 ± 6 cm/s. Relative to the calculated mean velocity for all patients, the standard deviation for the PSV at 60° was nearly twice that recorded at 50° (7.9 versus 3.9). The best correlation of the calculated mean velocity for all patients existed between the angles of 45° and 50° [r(36) = 0.92; P < .001 for center-steered data; and r(40) = 0.96; P < .001 for right-steered data]. CONCLUSIONS: These results indicate a statistically significant difference in the PSV measurements taken at varying Doppler angles. The greatest mean, variance, and lowest correlations all result when using 60°. The findings support the need for consistent ultrasound techniques and suggest that further study is warranted regarding the optimal Doppler angle for velocity measurements.
No takes yet. Share an insight, caveat, or question.
Campbell et al. (2020) conducted a cross-sectional in Asymptomatic (n=22). Varying angles of insonation (45°, 50°, 55°, and 60°) was evaluated on Peak systolic velocity (PSV) measurements (p=<.01). Varying the angle of insonation during internal carotid artery Doppler ultrasound significantly altered peak systolic velocity measurements (P < .01), with the greatest variance at 60°.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: