Answers are grounded in this paper and cite their sources.
The concentration and total injected volume of polymer solution affect elution volumes in size exclusion chromatography. When selecting optimal experimental conditions for both analytical and preparative separation, it must be considered wheter it is more advantageous to inject a smaller volume of solution of higher concentration, or vice versa. Theoretical analysis by means of derived equations and experimental results showed that if the injection conditions are chosen so that the contribution to the total width of the elution curve due to injection width is negligible within the limits of experimental error, and the total amount of the injected polymer is constant, the ratio of concentration to the injected solution volume may be chosen within broad limits without affecting the results of separation to any considerable extent.
No takes yet. Share an insight, caveat, or question.
Josef Janča (1981) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: