Key result
Inspiron sirolimus-eluting stents reduce six-month in-segment late loss ~67% versus bare-metal Cronus stents.
Why the study?
The safety and efficacy of the Inspiron sirolimus-eluting stent with abluminal biodegradable polymer and thin cobalt-chromium alloy had not been evaluated in humans.
Does the Inspiron sirolimus-eluting stent reduce in-segment late loss compared to a bare metal stent in patients requiring coronary stenting?
RCT (n=58)
randomised
Yes
Does the Inspiron sirolimus-eluting stent reduce in-segment late loss compared to a bare metal stent in patients requiring coronary stenting?
Absolute Event Rate: 0.19% vs 0.58%
p-value: p=<0.001
The novel Inspiron sirolimus-eluting stent with an abluminal biodegradable polymer significantly reduced in-segment late loss and neointimal obstruction at 6 months compared to a bare metal stent.
Inspiron reduced 6-month late loss versus bare-metal stent; leaves open confirmation of clinical benefit in larger trials.
AIMS: The INSPIRON-I trial is a first-in-man evaluation of the safety and efficacy of the Inspiron drug-eluting stent, a sirolimus-eluting stent with abluminal biodegradable polymer coating and thin cobalt-chromium alloy. METHODS AND RESULTS: This is a randomised, multicentre comparison between Inspiron and a stent with the same metallic structure but without polymer coating or drug elution (Cronus). The primary objective was to evaluate the in-segment late loss (LLL) at six months. Secondary endpoints included percent in-stent obstruction as measured by intravascular ultrasound (IVUS) at six months and major adverse cardiac events (MACE). Fifty-eight patients were enrolled (60 lesions), 39 for Inspiron and 19 for Cronus. Baseline clinical and angiographic characteristics of both groups were similar. At six months, the in-segment LLL was reduced in the Inspiron group compared to the control group (0.19±0.16 mm vs. 0.58±0.4 mm, respectively; p<0.001), as well as the percent neointimal obstruction (7.8±7.1% vs. 26.5±11.4%; p<0.001). At two-year follow-up, incidence of MACE was similar between groups (7.9 vs. 21.1%, respectively; p=0.20), with lower target lesion revascularisation for Inspiron (0 vs. 21.1%, respectively; p=0.01) and no stent thrombosis. CONCLUSIONS: Sirolimus eluted from an abluminal biodegradable polymer on a cobalt-chromium alloy proved effective in reducing restenosis at six months.
No takes yet. Share an insight, caveat, or question.
Ribeiro et al. (2014) conducted an RCT in Coronary artery disease (n=58). Inspiron drug-eluting stent vs. Cronus bare-metal stent was evaluated on in-segment late loss (LLL) at six months (p=<0.001). The Inspiron sirolimus-eluting stent significantly reduced in-segment late loss at six months compared to the bare-metal Cronus stent (0.19 mm vs. 0.58 mm; p<0.001).