Key result
The Svelte sirolimus-eluting coronary stent on an integrated delivery system demonstrated a 7% rate of target vessel failure and 0.22 mm in-stent late lumen loss at six months.
Why the study?
Does the Svelte sirolimus-eluting coronary stent on an integrated delivery system prevent target vessel failure and reduce late lumen loss in patients with de novo coronary artery lesions?
Does the Svelte sirolimus-eluting coronary stent on an integrated delivery system prevent target vessel failure and reduce late lumen loss in patients with de novo coronary artery lesions?
The first-in-human evaluation of the Svelte sirolimus-eluting coronary stent on an integrated delivery system demonstrated safety and efficacy with minimal in-stent proliferation at six months.
AIMS: The DIRECT study is a first-in-human evaluation of the safety and efficacy of the Svelte sirolimus-eluting coronary stent mounted on a fixed-wire, "all-in-one" integrated delivery system (IDS) in patients with de novo coronary artery lesions. The system permits easy delivery, deployment and post-dilatation of a cobalt-chromium stent eluting sirolimus from a fully bioabsorbable amino acid coating. The stent on its IDS has a very low profile, and is designed specifically to facilitate direct stenting. METHODS AND RESULTS: Patients with symptomatic ischaemic heart disease and a single de novo native coronary lesion suitable for percutaneous coronary intervention were prospectively enrolled at four New Zealand sites. The lesion length had to be <23 mm and the vessel reference diameter 2.5-3.5 mm. The primary safety and efficacy endpoints were target vessel failure (TVF) and angiographic in-stent late lumen loss (LLL) at six months, respectively. Twenty-nine of 30 enrolled patients completed six-month follow-up. TVF occurred in two patients (7%). The in-stent LLL was 0.22 ± 0.27 mm. No patient had clinically-driven target lesion revascularisation. Intravascular ultrasound neointimal volume was 3.3 ± 4.4 mm3 and volume obstruction was 2.7 ± 4.5% at six months. Optical coherence tomography showed 98 ± 4% strut coverage at a depth of 0.12 ± 0.06 mm. No patient developed stent thrombosis. CONCLUSIONS: Percutaneous coronary intervention using the Svelte sirolimus-eluting coronary stent mounted on an IDS appears safe and effective in de novo coronary artery lesions, with minimal in-stent proliferation and excellent stent strut coverage at six months.
No takes yet. Share an insight, caveat, or question.
Webster et al. (2013) studied symptomatic ischaemic heart disease and a single de novo native coronary lesion (n=30). Svelte sirolimus-eluting coronary stent on an integrated delivery system was evaluated on target vessel failure (TVF) and angiographic in-stent late lumen loss (LLL) at six months. The Svelte sirolimus-eluting coronary stent on an integrated delivery system demonstrated a 7% rate of target vessel failure and 0.22 mm in-stent late lumen loss at six months.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: