Key result
Implantation of the Svelte stent IDS was safe and feasible, with a 97.9% lesion success rate and 97.9% freedom from major adverse cardiac events at 30 days.
Why the study?
Is the Svelte Stent Integrated Delivery System safe and feasible in patients undergoing coronary stenting?
Is the Svelte Stent Integrated Delivery System safe and feasible in patients undergoing coronary stenting?
The first-in-man study of the Svelte stent IDS demonstrated safety and feasibility with low acute vascular injury and high strut coverage at 6 months, though with a 23.4% overall TLR rate.
Supports Svelte IDS safety in stenting; hypothesis-generating and requires randomized trials before practice change.
AIMS: The Svelte Stent Integrated Delivery System (IDS) is a novel fixed-wire thin-strut cobalt-chromium stent characterised by a very low entry profile. The aim of the present study is to evaluate the safety and the feasibility of the Svelte stent IDS implantation in humans. METHODS AND RESULTS: The present investigation is a prospective, multicentre non-randomised single-arm study. The primary endpoint was freedom from major adverse cardiac events (MACE) at 30 days post-procedure. Invasive follow-up was scheduled at six months post implantation. A total of 47 patients were enrolled and serial OCT imaging was performed in a subgroup of 18 patients. At the index procedure the lesion success rate was 97.9% (46 patients), the mean acute gain was 1.56 ± 0.43 mm with a mean minimum lumen diameter of 2.48 ± 0.43 mm. Post-implantation OCT imaging revealed a minimal mean prolapse area (0.10 ± 0.06 mm²), mean incomplete stent apposition area (0.12 ± 0.13 mm²) and mean intraluminal mass area (0.05 ± 0.03 mm²). Edge dissections were reported in eight cases (mean dissection width 0.17 ± 0.07 mm proximally and 0.25 ± 0.24 mm distally). At 30-day clinical follow-up, one case of myocardial infarction was reported. At six months, the angiographic mean in-stent late loss was 0.95 ± 0.76 mm. By OCT, a high percentage of struts was covered (97.6 ± 15.00 %) with a mean neointimal thickness of 0.31 ± 0.14 mm, all edge dissections were clinically silent and healed. Target lesion revascularisation (TLR) occurred in 11 patients (23.4%) and clinically driven TLR in three of these patients (6.4%). No cases of death or stent thrombosis were reported during the study. CONCLUSIONS: Implantation of the Svelte stent IDS was observed to be safe, feasible and associated with a low acute vascular injury and a high percentage of strut coverage at 6-month follow-up.
No takes yet. Share an insight, caveat, or question.
Diletti et al. (2013) studied Coronary artery disease (n=47). Svelte Stent Integrated Delivery System (IDS) was evaluated on freedom from major adverse cardiac events (MACE) at 30 days post-procedure. Implantation of the Svelte stent IDS was safe and feasible, with a 97.9% lesion success rate and 97.9% freedom from major adverse cardiac events at 30 days.