Key result
Arrhythmic event analysis detects ~76% of CIED lead failures, outperforming conventional impedance abnormalities.
Why the study?
Lead failure in patients with cardiovascular implantable electronic devices is often identified only by arrhythmic events and not by conventional impedance abnormalities, creating a need to compare these detection methods during remote monitoring.
Does analysis of arrhythmic events improve the detection of lead failure compared to conventional impedance abnormalities in patients with CIEDs followed by remote monitoring?
Observational (n=1,873)
Yes
Does analysis of arrhythmic events improve the detection of lead failure compared to conventional impedance abnormalities in patients with CIEDs followed by remote monitoring?
Careful human analysis of arrhythmic events during remote monitoring is crucial for early detection of CIED lead failure, as it identifies significantly more failures than conventional impedance alerts.
Answers are grounded in this paper and cite their sources.
May identify more CIED lead failures than impedance monitoring; hypothesis-generating and requires prospective validation.
BACKGROUND: Remote monitoring (RM) has been advocated as the new standard of care for patients with cardiovascular implantable electronic devices (CIEDs). RM has allowed the early detection of adverse clinical events, such as arrhythmia, lead failure, and battery depletion. However, lead failure was often identified only by arrhythmic events, but not impedance abnormalities. OBJECTIVE: To compare the usefulness of arrhythmic events with conventional impedance abnormalities for identifying lead failure in CIED patients followed by RM. METHODS: CIED patients in 12 hospitals have been followed by the RM center in Okayama University Hospital. All transmitted data have been analyzed and summarized. RESULTS: From April 2009 to March 2016, 1,873 patients have been followed by the RM center. During the mean follow-up period of 775 days, 42 lead failure events (atrial lead 22, right ventricular pacemaker lead 5, implantable cardioverter defibrillator [ICD] lead 15) were detected. The proportion of lead failures detected only by arrhythmic events, which were not detected by conventional impedance abnormalities, was significantly higher than that detected by impedance abnormalities (arrhythmic event 76.2%, 95% CI: 60.5-87.9%; impedance abnormalities 23.8%, 95% CI: 12.1-39.5%). Twenty-seven events (64.7%) were detected without any alert. Of 15 patients with ICD lead failure, none has experienced inappropriate therapy. CONCLUSIONS: RM can detect lead failure earlier, before clinical adverse events. However, CIEDs often diagnose lead failure as just arrhythmic events without any warning. Thus, to detect lead failure earlier, careful human analysis of arrhythmic events is useful.
No takes yet. Share an insight, caveat, or question.
Nishii et al. (2017) conducted an observational in Cardiovascular implantable electronic devices (CIEDs) (n=1,873). Analysis of arrhythmic events vs. Conventional impedance abnormalities was evaluated on Proportion of lead failures detected. Analysis of arrhythmic events detected a higher proportion of CIED lead failures than conventional impedance abnormalities.