Key result
Angiolite SES proves noninferior to EES for late lumen loss at 6 months.
Why the study?
The Angiolite durable fluoroacrylate polymer-based sirolimus-eluting stent has shown promising preclinical and clinical results, but its performance compared to an everolimus-eluting stent in coronary artery disease patients was not established.
Does a durable fluoroacrylate polymer-based sirolimus-eluting stent reduce late lumen loss compared to an everolimus-eluting stent in patients with coronary artery disease?
RCT (n=223)
1:1
Yes
Does a durable fluoroacrylate polymer-based sirolimus-eluting stent reduce late lumen loss compared to an everolimus-eluting stent in patients with coronary artery disease?
The Angiolite sirolimus-eluting stent demonstrated non-inferior angiographic late lumen loss at 6 months compared to an everolimus-eluting stent, with similar clinical outcomes at 2 years.
Answers are grounded in this paper and cite their sources.
Noninferior late lumen loss supports Angiolite as an alternative; confirms durable polymer sirolimus-eluting stents match everolimus-eluting performance in CAD.
AIMS: The durable fluoroacrylate polymer-based sirolimus-eluting stent (Angiolite SES) has shown promising preclinical and clinical results regarding inflammatory vascular reaction and neointimal healing. We aimed to compare performance between the Angiolite SES and an everolimus-eluting stent (EES) in patients with coronary artery disease. METHODS AND RESULTS: The ANGIOLITE trial, a prospective, randomised, multicentre trial, compared the restenosis parameters of both stents in de novo coronary lesions. The primary endpoint was late lumen loss at six-month angiographic follow-up. In-stent healing was assessed by optical coherence tomography (OCT). The main clinical endpoint was target lesion failure (TLF) evaluated up to 24 months. A total of 223 patients were randomised 1:1 to EES or SES. At six months, in-stent late lumen loss was 0.08 mm (±0.38) for EES vs 0.04 mm (±0.39) for SES (difference=-0.04 mm, 95% CI: -0.15, 0.07, p for non-inferiority=0.002). By OCT, the rate of uncovered to total number of struts score >30% was comparable between the groups whereas neointimal thickness was reduced in the SES arm (9.0% [7.6, 10.6] vs 9.9% [8.5, 11.3], p=0.41; and 86.4 [81.6, 91.2] µm vs 72.1 [68.2, 76.0] µm, p<0.01, respectively). At 24 months, TLF occurred in eight patients (7.6% [3.3, 14.5]) in the EES arm and in seven patients (7.1% [2.9, 14.0]) in the SES arm (p=0.88). The definite/probable stent thrombosis rate was comparable between the groups (1.9% [0.2, 6.7] vs 1.0% [0.0, 5.5] EES vs SES, respectively; p=0.59). CONCLUSIONS: This trial demonstrates similar antirestenotic efficacy at midterm follow-up of the Angiolite SES vs an EES. Clinical endpoints were comparable between the groups at two-year follow-up. Visual summary. Main results of the ANGIOLITE trial.
No takes yet. Share an insight, caveat, or question.
Moreu et al. (2019) conducted an RCT in Coronary artery disease (n=223). Angiolite durable fluoroacrylate polymer-based sirolimus-eluting stent (SES) vs. Durable fluoropolymer-based everolimus-eluting stent (EES) was evaluated on Late lumen loss at six-month angiographic follow-up (MD -0.04 mm, 95% CI -0.15, 0.07, p=0.002 for non-inferiority). The Angiolite sirolimus-eluting stent was noninferior to an everolimus-eluting stent for late lumen loss at 6 months.