Answers are grounded in this paper and cite their sources.
In this paper, a piezoelectric (PZ) sensor based on quartz crystal microbalance (QCM) for the determination of gluconic acid is described. PZ quartz crystals were coated with a 550-700 nm-thick layer of nanoflbers comprising of poly(DL-lactide-co-glycolide) (PLGA) and fullerene-C60 by electrospinning. Glucose oxidase (GOx) was immobilized on coverslips electrospun with nanofibrous PLGA for the interaction of the PZ sensor with gluconic acid, the oxidation product of β-D-glucose. Immobilization of GOx was confirmed by the activity assay and scanning electron microscopy analyses. The interaction of fullerene-C60 with electron-withdrawing gluconic acid molecules was followed by the change in the resonance frequency of the coated crystals in a flow system. The effects of glucose (gluconic acid) concentration, interfering species and temperature on the frequency response of the PLGA/C60-QCM sensor was investigated. The resonance frequency decreased owing to the adsorption of gluconic acid on the sensor. A linear relationship between the frequency response and gluconic acid concentration was observed in the range of 1.4-14.0 mM at room temperature. The optimal temperature of 37°C for gluconic acid formation was also shown by frequencies recorded with the PZ sensor. A good selectivity for gluconic acid, in the presence of some potential interferents, namely urea, ascorbic acid, cysteine, ammonia, Na+, K+, creatine, acetone, and ethanol was also determined.
No takes yet. Share an insight, caveat, or question.
Şeker et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: