Key result
Adjustable mitral ring reduces EROA ~80% on the beating heart in a swine model.
Why the study?
Residual mitral regurgitation after valve repair worsens clinical outcomes and postimplant adjustable mitral rings may allow reshaping of the annulus on the beating heart to correct it.
Does a postimplant adjustable mitral ring reduce residual mitral regurgitation in a swine model of severe MR?
Does a postimplant adjustable mitral ring reduce residual mitral regurgitation in a swine model of severe MR?
Answers are grounded in this paper and cite their sources.
A novel postimplant adjustable mitral ring successfully and reversibly corrected severe mitral regurgitation on the beating heart in a swine model.
OBJECTIVES: Residual mitral regurgitation after valve repair worsens patients' clinical outcome. Postimplant adjustable mitral rings potentially address this issue, allowing the reshaping of the annulus on the beating heart under echocardiography control. We developed an original mitral ring allowing valve geometry remodelling after the implantation and designed an animal study to assess device effectiveness in correcting residual mitral regurgitation. METHODS: The device consists of two concentric rings: one internal and flexible, sutured to the mitral annulus and a second external and rigid. A third conic element slides between the two rings, modifying the shape of the flexible ring. This sliding element is remotely activated with a rotating tool. Animal model: in adult swine, under cardio pulmonary bypass and cardiac arrest, we shortened the primary chordae of P2 segment to reproduce Type III regurgitation and implanted the active ring. We used intracardiac ultrasound to assess mitral regurgitation and the efficacy of the active ring to correct it. RESULTS: Severe mitral regurgitation (3+ and 4+) was induced in eight animals, 54 ± 6 kg in weight. Vena contracta width decreased from 0.8 ± 0.2 to 0.1 cm; proximal isovelocity surface area radius decreased from 0.8 ± 0.2 to 0.1 cm and effective regurgitant orifice area decreased from 0.50 ± 0.1 to 0.1 ± 0.1 cm(2). Six animals had a reversal of systolic pulmonary flow that normalized following the activation of the device. All corrections were reversible. CONCLUSIONS: Postimplant adjustable mitral ring corrects severe mitral regurgitation through the reversible modification of the annulus geometry on the beating heart. It addresses the frequent and morbid issue of recurrent mitral valve regurgitation.
No takes yet. Share an insight, caveat, or question.
Tozzi et al. (2013) studied Residual mitral regurgitation (n=8). Postimplant adjustable mitral ring vs. Pre-activation baseline was evaluated on Mitral regurgitation severity (vena contracta width, PISA radius, effective regurgitant orifice area). In a swine model, a postimplant adjustable mitral ring corrected severe mitral regurgitation on the beating heart, reducing effective regurgitant orifice area from 0.50 ± 0.1 to 0.1 ± 0.1 cm2.
Synapse has enriched 10 related papers. Consider them for comparative context: