Key result
Novel adjustable mitral annuloplasty ring increases maximum coaptation height 7 mm in ex vivo swine.
Why the study?
Residual and recurrent mitral regurgitations may be corrected by improving mitral leaflets coaptation with a novel adjustable mitral annuloplasty ring on the beating heart.
Does a novel adjustable mitral annuloplasty ring improve mitral leaflets coaptation height in an ex vivo swine heart model?
Does a novel adjustable mitral annuloplasty ring improve mitral leaflets coaptation height in an ex vivo swine heart model?
A novel adjustable mitral annuloplasty ring demonstrated feasibility in increasing mitral leaflet coaptation height in an ex vivo swine beating heart model.
Answers are grounded in this paper and cite their sources.
Supports further device development; hypothesis-generating for mitral repair, no practice change yet.
This investigation sought to determine the feasibility of a novel mitral ring designed to reshape mitral annulus on beating heart, after surgery. The mitral ring is intended to improve mitral leaflets coaptation to correct residual and recurrent mitral regurgitations. It could also provide progressive correction of mitral regurgitation. The device was tested in ex vivo beating heart model. The novel mitral ring is selectively deformable in P1, P2, and P3 segments using a dedicated angioplasty-type balloon. The deformation should increase leaflets coaptation, reducing distance between the two leaflets. It was implanted using standard surgical techniques. The mock loop is based on passive beating heart. Mitral valve (MV) functioning was evaluated in terms of leaflets coaptation height at P2 level using epicardial echocardiography. The test has been completed on eight swine hearts. Ring size was 30 mm. The balloons were inserted in the connecting line. Each segment of the posterior annulus was independently activated over three progressive positions. Balloon inflation pressures were between 15 and 21 bar. Maximum coaptation height increase was 7 mm. Mean pressure gradient across the MV was 1.7 ± 0.3 mm Hg after complete activation of the device. The device allowed significant increase in coaptation height at P2 level after adjustments at P1, P2, and P3. Results were consistent and reproducible. This feasibility study demonstrates the possibility to reshape the mitral annulus on beating heart to precisely increase MV leaflets coaptation height.
No takes yet. Share an insight, caveat, or question.
Tozzi et al. (2016) studied Mitral regurgitation (n=8). Adjustable mitral annuloplasty ring was evaluated on Leaflets coaptation height at P2 level. A novel adjustable mitral annuloplasty ring allowed a maximum coaptation height increase of 7 mm and a mean pressure gradient of 1.7 ± 0.3 mm Hg in an ex vivo swine heart model.