Key result
Tensiometric pacing lead sensor proves feasible as a hemodynamic sensor for detecting myocardial contraction.
Why the study?
The study aimed to measure the charge generated by the triboelectric effect in implanted cardiac leads to assess feasibility as a hemodynamic sensor for myocardial contraction detection.
Can a tensiometric pacing lead sensor based on triboelectricity feasibly detect myocardial contraction?
Can a tensiometric pacing lead sensor based on triboelectricity feasibly detect myocardial contraction?
A tensiometric stylet based on triboelectricity is a feasible and easily exchangeable hemodynamic sensor for detecting myocardial contraction during pacemaker procedures.
Answers are grounded in this paper and cite their sources.
Does not yet support clinical adoption; leaves open validation of integrated hemodynamic sensing in pacing leads.
Cardiac contractions bend the implanted cardiac lead body, extend and compress the lead conductors, their insulation and the inserted stylet. Magnitude of lead deflection depends on cardiac muscle contraction forces. The purpose of study was to measure the charge generated due to triboelectric effect between one of the lead conductors and the inserted stylet. The charge was measured by differential charge amplifier being connected to isolation amplifier and power supply. Sensor signal, ECG and intracardiac electrograms were acquired. Three models of custom designed leads were implanted in 8 sheep. Measurements were done in 18 patients undergoing pacemaker implantation and replacement procedures. Atrial and ventricular tensiometric signals were recorded in dual chamber and in single-lead VDD patients. Recordings in sinus rhythm at various AV intervals and in supraventricular tachycardia were done. In average, charge variation between 1 and 600 pC was measured. Tensiometric stylet could be feasible hemodynamic sensor for myocardial contraction detection. Its main advantage is that it is easily exchangeable and universal for all leads.
No takes yet. Share an insight, caveat, or question.
Tomašić et al. (2011) studied Pacemaker implantation and replacement (n=18). Tensiometric pacing lead sensor (tensiometric stylet) was evaluated on Charge variation due to triboelectric effect. A tensiometric pacing lead sensor successfully measured charge variations between 1 and 600 pC, demonstrating feasibility as a hemodynamic sensor for myocardial contraction detection.