Key result
LBBAP linked to far-field P-wave oversensing requiring recognition at implantation to prevent syncope.
Why the study?
Far-field P-wave oversensing is a recognized issue in His bundle pacing but its occurrence in left bundle branch area pacing and management strategies require clarification.
Case Report (n=1)
Far-field P-wave oversensing is a potential complication of left bundle branch area pacing that can lead to syncope and requires recognition and management.
Post-implant syncope in LBBAP may reflect far-field oversensing; leaves open optimal detection and correction strategies.
Key Teaching Points•Syncopal episodes after pacemaker implantation can indicate ventricular oversensing.•Far-field P-wave oversensing is a known issue in His bundle pacing, but can also occur in left bundle branch area pacing.•Oversensing should be recognized at implantation. If not corrected for during implantation, possible solutions for the problem are optimizing sensitivity, changing to an asynchronous or triggered pacing mode, or reintervention. •Syncopal episodes after pacemaker implantation can indicate ventricular oversensing.•Far-field P-wave oversensing is a known issue in His bundle pacing, but can also occur in left bundle branch area pacing.•Oversensing should be recognized at implantation. If not corrected for during implantation, possible solutions for the problem are optimizing sensitivity, changing to an asynchronous or triggered pacing mode, or reintervention.
No takes yet. Share an insight, caveat, or question.
Puyvelde et al. (2023) conducted a case report in Far-field P-wave oversensing after pacemaker implantation (n=1). Left bundle branch area pacing was evaluated. Far-field P-wave oversensing can occur in left bundle branch area pacing and should be recognized at implantation to prevent syncopal episodes.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: