Key result
Targeted ICD reprogramming using ventricular far-field signals resolves His-channel undersensing and avoids lead revision.
Why the study?
His-bundle pacing connected to the atrial port of a dual-chamber ICD may lead to sensing failure and unsafe pacing.
Does targeted ICD reprogramming using ventricular far-field signals resolve His-channel undersensing and avoid lead revision in a patient with a His-bundle pacing lead connected to the atrial port of a dual-chamber ICD?
Case Report (n=1)
No
Does targeted ICD reprogramming using ventricular far-field signals resolve His-channel undersensing and avoid lead revision in a patient with a His-bundle pacing lead connected to the atrial port of a dual-chamber ICD?
Targeted device reprogramming utilizing ventricular far-field signals can successfully resolve His-bundle lead undersensing in a dual-chamber ICD, avoiding the need for invasive lead revision or reimplantation.
BACKGROUND: His-bundle pacing connected to the atrial port of a dual-chamber ICD may lead to sensing failure and unsafe pacing. CASE SUMMARY: A 67-year-old man with permanent atrial fibrillation, alternating rapid and slow ventricular rates despite medication, symptomatic pauses longer than 6s, and a dual chamber ICD (His lead in the RA port; DF4 lead in the RV port) presented with palpitations. ECG and 24-hour Holter demonstrated intermittent His-channel undersensing with pacing on the T-wave peak. Interrogation showed small His electrograms followed by a prominent, constant ventricular far-field signal on the atrial channel. We turned off far-field protection after V-sense, intentionally used the ventricular far-field signal for sensing of His lead, and disabled related discrimination algorithms. CONCLUSION: A dual-chamber ICD with a His-bundle pacing lead may be associated with complex programming challenges, and targeted reprogramming can avoid lead revision or reimplantation.
No takes yet. Share an insight, caveat, or question.
Vo et al. (2026) conducted a case report in Permanent atrial fibrillation with His-channel undersensing in a dual-chamber ICD (n=1). Targeted ICD reprogramming (turning off far-field protection after V-sense, using ventricular far-field signal for His sensing, and disabling discrimination algorithms) was evaluated on Resolution of His-channel undersensing and avoidance of lead revision or reimplantation. Targeted ICD reprogramming using the ventricular far-field signal resolved intermittent His-channel undersensing in 1 patient (67-year-old man) with pauses >6s, avoiding lead revision.