Key result
Vector flow imaging detects ~63% larger complex carotid flow patterns than color Doppler.
Why the study?
The effectiveness of high–frame rate vector flow imaging compared with color Doppler imaging for assessing complex flow patterns at the carotid bifurcation in healthy adults was not established.
Does high-frame rate vector flow imaging improve the assessment of complex flow patterns in the carotid bifurcation of healthy adults compared to color Doppler imaging?
Cross-Sectional (n=60)
Does high-frame rate vector flow imaging improve the assessment of complex flow patterns in the carotid bifurcation of healthy adults compared to color Doppler imaging?
High-frame rate vector flow imaging is more powerful than color Doppler imaging in assessing the extension and duration of complex flow patterns in the carotid bifurcation.
Answers are grounded in this paper and cite their sources.
May enhance complex flow detection in research settings; leaves open clinical adoption over color Doppler in healthy adults.
Objectives To evaluate the carotid bifurcation in healthy adults using a commercial system equipped with high–frame rate vector flow imaging (VFI) based on the plane wave and to compare VFI with color Doppler imaging. Methods Carotid bifurcation diameters and flow characteristics of 60 vessels in 60 healthy volunteers were evaluated quantitatively and qualitatively to assess complex flow patterns and their extension and duration. Results Complex flow in the internal carotid artery (ICA) was associated with a statistically significant difference in the ΔICA sinus‐to‐common carotid artery (CCA) diameter ratio (the relative change in diameter between the CCA and ICA sinus.) Vector flow imaging and color Doppler imaging were in accordance when detecting complex flow in 96.7% of cases; in 3.3% of cases, only VFI identified small recirculation areas of short duration. Vector flow imaging highlighted a larger extension of the complex flow (mean ± SD, 47.7 ± 28.5 mm2; median, 45.5 mm2) compared with color Doppler imaging (mean, 29.2 ± 19.9 mm2; median, 29.5 mm2) and better depicted different complex flow patterns; a strong correlation (r = 0.84) was found between the ΔICA sinus‐to‐CCA diameter ratio and the complex flow extension. Vector flow imaging showed a longer duration of the flow disturbances (mean, 380 ± 218 milliseconds; median, 352.5 milliseconds) compared with color Doppler imaging (mean, 325 ± 206 milliseconds; median, 333 milliseconds), and there was a strong correlation (r = 0.92). Conclusions Vector flow imaging is as effective as color Doppler imaging in the detection of flow disturbances, but it is more powerful in the assessment of complex flow patterns.
No takes yet. Share an insight, caveat, or question.
Goddi et al. (2018) conducted a cross-sectional in Healthy adults (n=60). High-frame rate vector flow imaging vs. Color Doppler imaging was evaluated on Complex flow patterns, extension, and duration. Vector flow imaging detected a larger extension and longer duration (380 vs 325 ms) of complex flow patterns in the carotid bifurcation compared with color Doppler imaging.