Key result
Updated Micra delivery catheter requires ~20% more force to perforate ventricular tissue than the original.
Why the study?
Pericardial effusion due to delivery catheter perforation is a rare complication of leadless pacemaker implantation, motivating evaluation of an updated Micra delivery catheter to reduce perforation risk.
Does an updated Micra delivery catheter reduce the risk of cardiac perforation compared to the original catheter in preclinical models?
Does an updated Micra delivery catheter reduce the risk of cardiac perforation compared to the original catheter in preclinical models?
An updated Micra delivery catheter with increased surface area and a rounded tip significantly improves preclinical perforation performance, suggesting a potential reduction in clinical perforation risk.
Answers are grounded in this paper and cite their sources.
Supports clinical evaluation of the updated Micra catheter; leaves open whether perforation risk decreases in patients.
Background Leadless pacemakers have been developed to avoid some of the complications that are associated transvenous pacemakers. Pericardial effusion is a rare complication of leadless pacemaker implantation, which may result from perforation of the delivery catheter. In this study, we describe preclinical perforation performance of an updated Micra delivery catheter. Methods To assess preclinical perforation performance of the updated delivery catheter, three analyses were performed. First, Finite Element Analysis (FEA) computational modeling was performed to estimate the target tissue stress during Micra delivery catheter tenting. Second, benchtop perforation forces of ovine tissue were recorded for the original and updated delivery catheters. Finally, a Monte‐Carlo simulation combining human cadaveric Micra implant forces and human ventricular tissue perforation properties was performed to estimate clinical perforation performance. Results FEA modeling demonstrated a 66% reduction in target tissue stress when using the updated Micra delivery catheter (6.2 vs. 2.2 psi, Original vs. Updated Micra delivery catheter). Updated Micra delivery catheters required 20% more force to perforate porcine ventricular tissues in benchtop testing (μupd = 26.9N vs. μorg = 22.4N, p = .01). Monte‐Carlo Simulation of catheter performance in human cadaveric tissues predicts 28.5% reduction of catheter‐perforated cases with the updated delivery catheter. Conclusions This study, using computer modelling and benchtop experimentation, has indicated that increased surface area and rounding of the updated Micra catheter tip significantly improves preclinical perforation performance. It will be important to evaluate the impact of these catheter design changes with robust registry data.
No takes yet. Share an insight, caveat, or question.
Roberts et al. (2023) studied Leadless pacemaker implantation. Updated Micra delivery catheter vs. Original Micra delivery catheter was evaluated on Benchtop perforation forces of ventricular tissue (p=.01). The updated Micra delivery catheter required more force to perforate ventricular tissues than the original catheter.