Careful review of the right anterior oblique view before leadless pacemaker deployment is recommended to ensure the device is oriented away from the border between the right ventricular septum and the free wall to prevent perforation.
Answers are grounded in this paper and cite their sources.
Subclinical perforation risk after leadless pacemaker placement may warrant imaging vigilance; single case leaves open incidence and prevention trials.
A subclinical cardiac perforation by a device cup of the Micra™ transcatheter pacing system was suspected in a 78-year-old woman. During the procedure, the device cup was placed on the septum. The contrast media was injected before device deployment and remained outside of the myocardium. Later, a cardiac computed tomography scan visualized a protruded diverticular structure on the right ventricle. The contrast material remained in a pouch within the pericardium. To ensure the device is oriented away from the border between the right ventricular septum and the free wall, right anterior oblique view should be carefully reviewed before deployment.