Key result
Emergency aortotomy successfully retrieves an embolized mechanical mitral valve leaflet from the aortic bifurcation.
Why the study?
Escape of the prosthetic heart valve disc is a cause of prosthetic dysfunction requiring emergency surgery, with risks from embolized disc extrusion.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Structural failure of a Tri-technologies bileaflet mechanical valve can lead to leaflet embolization, requiring emergency valve replacement and subsequent retrieval of the escaped leaflet.
The escape of the prosthetic heart valve disc is one of the causes of prosthetic dysfunction that requires emergency surgery. The removal of the embolized disc should be carried out because of the risk of a progressive extrusion on the aortic wall. Several imaging techniques can be used for the detection of the missing disc localization. In this report we describe a 32-year-old man who underwent mitral valve replacement with a Tri-technologies bileaflet valve three years ago, and was admitted in cardiogenic shock. Transesophageal echocardiography showed acute-onset massive mitral regurgitation. The patient underwent emergency replacement of the prosthetic valve. Only one of the two leaflets remained in the removed prosthetic valve. The missing leaflet could not be found within the cardiac cavity. The abdominal fluoroscopic study and plain radiography were unable to detect the escaped leaflet. The abdominal computed tomography scan and the ultrasound showed the escaped leaflet in the terminal portion of the aortic bifurcation. To retrieve the embolized disc laparotomy and aortotomy were performed three months later. The escaped leaflet shows a fracture of one of the pivot systems caused by structural failure. This kind of failure mode is usually the result of high stress concentration.
No takes yet. Share an insight, caveat, or question.
CIANCIULLI et al. (2006) conducted a case report in Prosthetic valve dysfunction (escaped leaflet) (n=1). Emergency valve replacement and leaflet retrieval was evaluated on Retrieval of the embolized disc. An escaped leaflet from a Tri-technologies bileaflet mechanical mitral valve was successfully located in the aortic bifurcation and retrieved via laparotomy and aortotomy.
Synapse has enriched 10 related papers. Consider them for comparative context: