Answers are grounded in this paper and cite their sources.
In this paper, experiences and learnings are shared from the 10-year application of incurred sample reanalysis (ISR) in support of the AstraZeneca small molecule portfolio. The conclusions from including ISR in every clinical bioanalysis study for a period of 5 years, generating ISR data from 550 studies, are shared. Our preclinical ISR approach is described and data generated using capillary microsampling demonstrate confidence in its routine application. The data demonstrate that ISR failures are very rare and the assessment can and should therefore be limited. Dialogue between the bioanalytical teams internally, as well as with the partner contract research organizations, is however critical for a successful bioanalytical method validation and to avoid any ISR failures.
No takes yet. Share an insight, caveat, or question.
Arfvidsson et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: