Key result
Prosthesis-anterior mitral leaflet interference post-TAVR is linked to ~2.3 mm shorter mitral-aortic annulus distance.
Why the study?
Scarce data exist regarding the occurrence and clinical impact of mitral valve interference after TAVR with Venus-A valve implantation.
What anatomical factors are associated with prosthesis-anterior mitral leaflet interference after TAVR with the Venus-A valve, and does it affect clinical outcomes?
Observational (n=20)
No
What anatomical factors are associated with prosthesis-anterior mitral leaflet interference after TAVR with the Venus-A valve, and does it affect clinical outcomes?
Interference between the Venus-A valve and the anterior mitral leaflet after TAVR is driven by specific anatomical and procedural factors like deeper implantation and shorter mitral-aortic annulus distance, leading to increased early perivalvular leakage but no significant difference in 1-year adverse events.
Answers are grounded in this paper and cite their sources.
AML interference after Venus-A TAVR links to deeper implantation and higher early PVL; leaves open long-term clinical relevance pending larger prospective data.
Background Scarce data exist regarding the occurrence of mitral valve interference after transcatheter aortic valve replacement (TAVR) with Venus-A valve implantation. Several case reports have noted that the anterior mitral leaflet (AML) is mechanically affected by the prosthesis frame, particularly when implanted in a low position. This study aimed to investigate the potential factors influencing the clinical outcomes of AML interference after Venus-A valve implantation. Methods We retrospectively included 20 severe aortic valve stenosis patients who had undergone TAVR and had been implanted with the Venus-A valve at our hospital between October 2020 and June 2021. Pre- and post-procedural CT scans were used for the FEops HEARTguide simulation. Anatomically influencing factors were measured using the 3mensio software and derived from the FEops HEARTguide. The prosthesis-AML interference (PAI) was defined when it met both of two criteria:1) significant interference and limited AML movement shown by transthoracic or transoesophageal echocardiography, and 2) more than half cell intersection between the simulated Venus-A valve and the reconstructed AML revealed by the FEops HEARTguide. Anatomical factors and clinical outcomes were compared between the PAI and non-PAI groups. Results Nine PAI patients and 11 non-PAI cases were identified. PAI was associated with shorter mitral-aortic annulus distance (2.7±1.7 mm vs 5.0±2.2 mm, P = 0.019), larger prosthesis valve size ( P = 0.013), deeper implantation (12.2±3.3 mm vs 6.2±2.9 mm at non-coronary cusp side, P < 0.001) and less calcification of non-coronary cusp (median calcification score, 52.2 mm 3 vs 156.0 mm 3 , P = 0.046). Regarding the clinical impact, PAI was associated with a higher rate of moderate or severe perivalvular leakage before discharge than those associated with the absence of PAI, with no difference in haemodynamic parameters and incidence of adverse events at the 30-day and 12-month follow-ups between the groups. Conclusions Interference between the Venus-A prosthesis valve and AML after TAVR was associated with a shorter mitral-aortic annulus distance, larger prosthesis usage, greater implantation depth, and less calcification of the non-coronary cusp. However, further studies are required to explore its long-term clinical impact.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2023) conducted an observational in Severe aortic valve stenosis (n=20). Prosthesis-anterior mitral leaflet interference (PAI) vs. No prosthesis-anterior mitral leaflet interference was evaluated on Mitral-aortic annulus distance (p=0.019). Prosthesis-anterior mitral leaflet interference after Venus-A valve implantation was associated with a shorter mitral-aortic annulus distance, larger prosthesis usage, and deeper implantation.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: