Background: Achieving stable fixation when performing tibiotalocalcaneal arthrodesis can be challenging, especially in osteopenic bone. The purpose of the current investigation was to compare the stiffness and fatigue endurance of blade-plate-and-screw fixation with intramedullary rod fixation in a cadaveric model. Methods: In ten matched pairs of fresh-frozen cadaveric legs, a tibiotalocalcaneal arthrodesis was performed with use of a blade-plate and a 6.5-mm sagittal screw in one leg and with use of an intramedullary rod in the contralateral leg. After an initial load-deformation curve was obtained, each specimen was loaded to 270 N through 250,000 cycles at a rate of 3 Hz. Results: Blade-plate-and-screw fixation resulted in significantly higher mean initial and final stiffness and decreased plastic deformation than did intramedullary rod fixation. In addition, there was an inverse correlation between bone-mineral density and the difference in plastic deformation noted between the specimens of each pair. Conclusions: Blade-plate fixation is biomechanically superior to intramedullary fixation for tibiotalocalcaneal arthrodesis. Clinical Relevance: Although good fixation can be achieved with an intramedullary rod or a blade-plate-and-screw construct, the stability of the blade-plate-and-screw construct may have advantages, particularly in osteopenic bone.
No takes yet. Share an insight, caveat, or question.
Chiodo et al. (2003) studied this question.