Purpose To develop a Compressed Sensing (CS)‐MP2RAGE sequence to drastically shorten acquisition duration and then detect and measure the T 1 of brain metastases in mice at 7 T. Methods The encoding trajectory of the standard Cartesian MP2RAGE sequence has been modified (1) to obtain a variable density Poisson disk under‐sampling distribution along the k y ‐k z plane, and (2) to sample the central part of the k‐space exactly at TI 1 and TI 2 inversion times. In a prospective study, the accuracy of the T 1 measurements was evaluated on phantoms containing increasing concentrations of gadolinium. The CS acceleration factors were increased to evaluate their influence on the contrast and T 1 measurements of brain metastases in vivo. Finally, the 3D T 1 maps were acquired with at 4‐fold increased spatial resolution. The volumes and T 1 values of the metastases were measured while using CS to reduce scan time. Results The implementation of the CS‐encoding trajectory did not affect the T 1 measurements in vitro. Accelerating the acquisition by a factor of 2 did not alter the contrast or the T 1 values of the brain metastases. 3D T 1 maps could be obtained in < 1 min using a CS factor of 6. Increasing the spatial resolution enabled more accurately measurement of the metastasis volumes while maintaining an acquisition duration below 5 min. Conclusion The CS‐MP2RAGE sequence could be of great interest in oncology to either rapidly obtain mouse brain 3D T 1 maps or to increase the spatial resolution with no penalty on the scan duration.
No takes yet. Share an insight, caveat, or question.
Trotier et al. (2018) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: