Key result
Brugada syndrome linked to ~266% higher S-ICD screening failure prevalence vs non-Brugada patients.
Why the study?
The prevalence of S-ICD ineligibility among conventional ICD recipients, especially comparing patients with and without Brugada syndrome, was unclear.
Does the presence of Brugada syndrome increase the rate of S-ICD screening failure in patients with conventional ICDs?
Cross-Sectional (n=130)
No
Does the presence of Brugada syndrome increase the rate of S-ICD screening failure in patients with conventional ICDs?
Patients with Brugada syndrome have a significantly higher rate of S-ICD screening failure, particularly those with low BMI, suggesting the need for careful screening and potential right parasternal electrode positioning.
Brugada syndrome was associated with higher S-ICD screening failure; hypothesis-generating and requires prospective validation before changing screening practice.
BACKGROUND: Subcutaneous implantable cardioverter-defibrillators (S-ICD) could eliminate lead-associated complications. We assessed the prevalence of S-ICD ineligibility in conventional ICD recipients and compared it in patients with and without Brugada syndrome (BrS). METHODS AND RESULTS: Consecutive patients with a transvenous ICD without an indication for antibradycardia pacing were assessed. A patient was considered eligible for S-ICD if the ECG satisfied the screening template, both supine and standing, in ≥1 lead. Among 130 patients (103 men, age 57±15 years), a total of 18 (13.8%) patients were ineligible. The BrS group (n=33) had a significantly higher prevalence of S-ICD screening failure as compared with the non-BrS group (P=0.003; 30% vs. 8.2%). In the BrS group, the body mass index (BMI) was significantly lower, and T/QRS amplitude in lead I was significantly higher in those who were ineligible than that in the patients who were eligible. Of the 10 BrS patients failing the screening, 4 became eligible in the right parasternal electrode position. CONCLUSIONS: Among current ICD patients, there was a high incidence of patients with BrS who were unsuitable for S-ICD based on the left parasternal screening test. Suitability screening of patients for S-ICDs should be conducted carefully in patients with BrS, particularly if the BMI is low. Right parasternal electrode positioning should also be tested in such BrS patients.
No takes yet. Share an insight, caveat, or question.
Kawabata et al. (2017) conducted a cross-sectional in Brugada syndrome in conventional ICD recipients (n=130). Brugada syndrome vs. Non-Brugada syndrome was evaluated on S-ICD screening failure (ineligibility) (p=0.003). Brugada syndrome was associated with a significantly higher prevalence of subcutaneous implantable cardioverter-defibrillator screening failure compared to non-Brugada patients (30% vs. 8.2%, P=0.003).