Key result
Among patients discharged with a wearable cardioverter-defibrillator after cardiac surgery, ventricular arrhythmias occurred in 13%, and mean LVEF improved from 28.9% to 36.7% (p<0.001).
Why the study?
Does a wearable cardioverter-defibrillator (WCD) protect against ventricular tachyarrhythmias and allow for LVEF recovery in patients following cardiac surgery?
Cohort (n=100)
No
Does a wearable cardioverter-defibrillator (WCD) protect against ventricular tachyarrhythmias and allow for LVEF recovery in patients following cardiac surgery?
WCDs effectively protect against ventricular tachyarrhythmias after cardiac surgery and allow time for LVEF recovery, potentially avoiding permanent ICD implantation.
Answers are grounded in this paper and cite their sources.
May support WCD bridging for LVEF recovery post-surgery; observational data leaves open need for RCTs versus observation alone.
Background A wearable cardioverter–defibrillator (WCD) can terminate ventricular fibrillation and ventricular tachycardias via electrical shock and thus give transient protection from sudden cardiac death. We investigated its role after cardiac surgery. Methods We retrospectively analyzed all patients who were discharged with a WCD from cardiac surgery department. The WCD was prescribed for patients with a left ventricular ejection fraction (LVEF) of ≤35% or an explanted implantable cardioverter–defibrillator (ICD). Results A total of 100 patients were included in this study, the majority (n = 59) had received coronary artery bypass graft surgery. The median wearing time of a WCD patient was 23.5 hours per day. LVEF was 28.9 ± 8% after surgery and improved in the follow-up to 36.7 ± 11% (p < 0.001). Three patients were successfully defibrillated. Ten patients experienced ventricular tachycardias. No inappropriate shocks were given. An ICD was implanted in 25 patients after the WCD wearing period. Conclusion Ventricular arrhythmias occurred in 13% of the investigated patients. LVEF improved significantly after 3 months, and thus a permanent ICD implantation was avoided in several cases. Sternotomy did not impair wearing time of the WCD. A WCD can effectively protect patients against ventricular tachyarrhythmias after cardiac surgery.
No takes yet. Share an insight, caveat, or question.
Heimeshoff et al. (2018) conducted a cohort in Post-cardiac surgery with LVEF ≤35% or explanted ICD (n=100). Wearable cardioverter-defibrillator was evaluated on Ventricular arrhythmias. Among patients discharged with a wearable cardioverter-defibrillator after cardiac surgery, ventricular arrhythmias occurred in 13%, and mean LVEF improved from 28.9% to 36.7% (p<0.001).