Key result
TEMEO telemonitoring achieves ~99% coincidence with standard ECGs and correlates highly with Holter parameters.
Why the study?
The clinical applicability of a telemetric heart rhythm monitoring system (TEMEO) compared to standard ECG methods needed evaluation.
Does the TEMEO telemonitoring system provide comparable ECG and Holter results to standard 12-lead ECG and Holter monitoring in patients?
Observational (n=60)
Does the TEMEO telemonitoring system provide comparable ECG and Holter results to standard 12-lead ECG and Holter monitoring in patients?
Answers are grounded in this paper and cite their sources.
The TEMEO mobile heart rhythm telemonitoring system provides highly comparable results to standard 12-lead ECG and Holter monitoring, supporting its clinical applicability.
Purpose. To evaluate the clinical applicability of a telemonitoring system: telemetric system for collection and distant surveillance of medical information (TEMEO). Methods. We evaluated 60 patients, applying simultaneously standard Holter ECG and telemonitoring. Two different comparisons were performed: (1) TEMEO ECG with standard 12-lead ECG; (2) TEMEO Holter with standard ECG Holter. Results. We found a very high coincidence rate (99.3%) between TEMEO derived ECGs and standard ECGs. Intraclass correlation coefficient analysis revealed high and significant correlation coefficients regarding average, maximal, and minimal heart rate, % of time in tachycardia, single supraventricular ectopic beats (SVEB), and single and couplets of ventricular ectopic beats (VEB) between Holter ECG and TEMEO derived parameters. Couplets of SVEB were recorded as different by the two monitoring systems, however, with a borderline statistical significance. Conclusions. TEMEO derived ECGs have a very high coincidence rate with standard ECGs. TEMEO patient monitoring provides results that are similar to those derived from a standard Holter ECG.
No takes yet. Share an insight, caveat, or question.
Mateev et al. (2012) reported an observational. TEMEO telemonitoring system vs. Standard 12-lead ECG and standard Holter ECG was evaluated on Coincidence rate between TEMEO derived ECGs and standard ECGs. The TEMEO telemonitoring system demonstrated a 99.3% coincidence rate with standard ECGs and showed high correlation with standard Holter ECG parameters.