Single-lead ECG patch telemonitoring cuts signal noise to with conventional monitoring.
Why the study?
The comparative effectiveness of a single-lead ECG patch for real-time telemonitoring in inpatient settings remained unclear.
Does a novel telemonitoring system using a single-lead ECG patch offer comparable performance to a conventional telemetry system for arrhythmia monitoring in hospitalized patients?
Cohort (n=80)
No
Does a novel telemonitoring system using a single-lead ECG patch offer comparable performance to a conventional telemetry system for arrhythmia monitoring in hospitalized patients?
Answers are grounded in this paper and cite their sources.
A novel single-lead ECG patch for inpatient telemetry provides excellent reliability for detecting arrhythmias while significantly reducing signal loss and noise compared to conventional multi-lead telemetry.
BACKGROUND AND OBJECTIVES: Although a single-lead electrocardiogram (ECG) patch may provide advantages for detecting arrhythmias in outpatient settings owing to user convenience, its comparative effectiveness for real-time telemonitoring in inpatient settings remains unclear. We aimed to compare a novel telemonitoring system using a single-lead ECG patch with a conventional telemonitoring system in an inpatient setting. METHODS: This was a single-center, prospective cohort study. Patients admitted to the cardiology unit for arrhythmia treatment who required a wireless ECG telemonitoring system were enrolled. A single-lead ECG patch and conventional telemetry were applied simultaneously in hospitalized patients for over 24 hours for real-time telemonitoring. The basic ECG parameters, arrhythmia episodes, and signal loss or noise were compared between the 2 systems. RESULTS: Eighty participants (mean age 62±10 years, 76.3% male) were enrolled. The three most common indications for ECG telemonitoring were atrial fibrillation (66.3%), sick sinus syndrome (12.5%), and atrioventricular block (10.0%). The intra-class correlation coefficients for detecting the number of total beats, atrial and ventricular premature complexes, maximal, average, and minimal heart rates, and pauses were all over 0.9 with p values for reliability <0.001. Compared to a conventional system, a novel system demonstrated significantly lower signal noise (median 0.3% [0.1-1.6%] vs. 2.4% [1.4-3.7%], p<0.001) and fewer episodes of signal loss (median 22 [2-53] vs. 64 [22-112] episodes, p=0.002). CONCLUSIONS: The novel telemonitoring system using a single-lead ECG patch offers performance comparable to that of a conventional system while significantly reducing signal loss and noise. TRIAL REGISTRATION: Clinical Research Information Service Identifier: KCT0008176.
No takes yet. Share an insight, caveat, or question.
Kwon et al. (2024) conducted a cohort in Cardiac arrhythmias requiring telemonitoring (n=80). Single-lead ECG patch telemonitoring system vs. Conventional 3-channel telemetry device was evaluated on Proportion of signal noise data (p=<0.001). The novel telemonitoring system using a single-lead ECG patch demonstrated excellent reliability for ECG parameters while reducing the proportion of signal noise.