Key result
Monopolar cautery with a full-body return electrode is linked to inappropriate ICD discharges during surgery.
Why the study?
Current consensus guidelines do not recommend device reprogramming or magnet use for surgeries below the umbilicus, but these guidelines did not address use of full-body return electrodes with electrocautery in patients with CIEDs.
Case Report (n=1)
The use of full-body return electrodes during subumbilical surgery with monopolar cautery can cause inappropriate ICD discharges, suggesting a need to update current consensus guidelines.
Full-body return electrodes may trigger inappropriate ICD therapies during monopolar cautery; hypothesis-generating for guideline updates in device patients.
Patients with cardiac implantable electronic devices (CIED) and implantable cardioverter-defibrillator (ICD) devices frequently present for surgical procedures. If electrocautery is used, careful planning is needed to avoid inappropriate device function or device damage. Published consensus statements suggest that if the surgery is below the umbilicus, interference is typically minimal, and therefore it is not recommended to reprogram or disable the CIED. When these guidelines were published, full-body return electrodes were not commonly used in clinical practice, and therefore were not addressed in the recommendations. A 76-year-old male with a single chamber ICD underwent bladder surgery under general anesthesia. Monopolar cautery was used with a full-body return electrode. The patient had undergone a similar procedure multiple times prior utilizing a traditional thigh adhesive return electrode without any inappropriate ICD discharges. During the procedure, the patient's movement was noted with electrocautery use which was suspected to be an inappropriate discharge of his ICD. Device interrogation was performed confirming two antitachycardia pacing therapies and four defibrillations due to interference from the electrocautery. This case examines inappropriate ICD discharge related to interference from electrocautery when utilizing a full-body return electrode, despite a subumbilical location of surgery. Current consensus statement guidelines do not recommend device reprogramming or magnet used when surgery is below the umbilicus, however, these full-body return electrodes were not routinely used when these guidelines were published. Based on these results, the authors avoid full-body return electrodes in patients with CIEDs.
No takes yet. Share an insight, caveat, or question.
Lefevre et al. (2020) conducted a case report in Patients with cardiac implantable electronic devices undergoing surgery (n=1). Monopolar cautery with a full-body return electrode vs. Traditional thigh adhesive return electrode (historical) was evaluated on Inappropriate ICD discharge. Use of monopolar cautery with a full-body return electrode during subumbilical surgery caused inappropriate ICD discharges, including two antitachycardia pacing therapies and four defibrillations.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: