Key result
Ex vivo SAPIEN 3 in 4 calcified CoreValve/Evolut explants yields adequate hemodynamics despite frame underexpansion.
Why the study?
Implantation of SAPIEN 3 valves for redo-TAVI in degenerated calcified CoreValve/Evolut transcatheter aortic valves is indicated but not well understood.
Does SAPIEN 3 implantation in calcified CoreValve/Evolut explants provide adequate hydrodynamic performance?
Does SAPIEN 3 implantation in calcified CoreValve/Evolut explants provide adequate hydrodynamic performance?
Ex vivo testing demonstrates that SAPIEN 3 implantation in degenerated, calcified CoreValve/Evolut valves yields satisfactory acute hydrodynamic performance, though frame underexpansion and leaflet pinwheeling occur.
Supports hydrodynamic feasibility in explants; hypothesis-generating and requires in vivo confirmation before clinical consideration.
BACKGROUND: Redo-transcatheter aortic valve implantation (TAVI) is the treatment of choice for failed transcatheter aortic valves. Currently, implantation of a SAPIEN 3 (S3) is indicated for redo-TAVI in degenerated CoreValve/Evolut (CV/EV) transcatheter aortic valves (TAVs) but is not well understood. AIMS: We aimed to evaluate S3 function following implantation in explanted calcified CV/EV TAVs and to assess the impact of CV/EV pathology on redo-TAVI outcomes. METHODS: Ex vivo hydrodynamic testing was performed per the International Organization for Standardization (ISO) 5840-3 standard on 4 S3 TAVs implanted at node 5 in calcified CV/EV explants. The mean gradient (MG), effective orifice area (EOA), peak velocity, regurgitant fraction (RF), geometric orifice area (GOA), leaflet overhang, leaflet pinwheeling, neoskirt height, and frame deformation were evaluated. RESULTS: CV/EV explants were calcified and stenotic. Following S3 implantation, the MG and peak velocity decreased. As per the ISO standard, all S3 implants showed adequate EOA, and 3 out of 4 had an RF within the accepted value (<20%). CV/EV leaflet overhang ranged from 25-37%. Calcified leaflets remained stationary throughout the cardiac cycle (difference <9%) and were not pinned in a manner that constrained S3 systolic flow or appeared to prevent selective frame cannulation. The downstream CV/EV GOA was larger than the upstream S3 GOA during systole. S3 frame underexpansion was seen, resulting in leaflet pinwheeling (range 13-30%). Above the neoskirt, calcium protrusion was observed in contact with the S3 leaflets. CONCLUSIONS: S3 implantation at node 5 in calcified CV/EV valves resulted in satisfactory hydrodynamic performance in most configurations tested with stable leaflet overhang throughout the cardiac cycle. The long-term implications of S3 underexpansion, leaflet pinwheeling, and calcium protrusion require future studies.
No takes yet. Share an insight, caveat, or question.
Meier et al. (2024) studied Degenerated calcified CoreValve/Evolut explants (n=4). SAPIEN 3 implantation was evaluated on Hydrodynamic performance (mean gradient, effective orifice area, peak velocity, regurgitant fraction). Ex vivo SAPIEN 3 implantation in 4 calcified CoreValve/Evolut explants yielded adequate effective orifice areas in all and acceptable regurgitant fraction (<20%) in 3, despite frame underexpansion.