Key result
Emergency leadless pacemaker implantation is not linked to increased major complications versus elective implantation.
Why the study?
Data on leadless pacemaker implantation in an emergency setting for urgent bradyarrhythmia were lacking.
Does leadless pacemaker implantation in an emergency setting increase peri-procedural complications compared to an elective setting in patients with bradyarrhythmia?
Cohort (n=1,154)
Yes
Does leadless pacemaker implantation in an emergency setting increase peri-procedural complications compared to an elective setting in patients with bradyarrhythmia?
Absolute Event Rate: 6.9% vs 4.2%
p-value: p=0.244
Leadless pacemaker implantation is feasible and safe for emergency bradyarrhythmia, with complication rates comparable to elective procedures despite longer procedural times.
Supports safety of emergency leadless pacemaker implantation in this cohort; leaves open need for randomized confirmation before practice change.
Background. Data on leadless pacemaker (LPM) implantation in an emergency setting are currently lacking. Objective. We aimed to investigate the feasibility of LPM implantation for emergency bradyarrhythmia, in patients referred for urgent PM implantation, in a large, multicenter, real-world cohort of LPM recipients. Methods. Two cohorts of LPM patients, stratified according to the LPM implantation scenario (patients admitted from the emergency department (ED+) vs. elective patients (ED−)) were retrieved from the iLEAPER registry. The primary outcome of the study was a comparison of the peri-procedural complications between the groups. The rates of peri-procedural characteristics (overall procedural and fluoroscopic duration) were deemed secondary outcomes. Results. A total of 1154 patients were enrolled in this project, with patients implanted due to an urgent bradyarrhythmia (ED+) representing 6.2% of the entire cohort. Slow atrial fibrillation and complete + advanced atrioventricular blocks were more frequent in the ED+ cohort (76.3% for ED+ vs. 49.7% for ED−, p = 0.025; 37.5% vs. 27.3%, p = 0.027, respectively). The overall procedural times were longer in the ED+ cohort (60 (45−80) mins vs. 50 (40−65) mins, p < 0.001), showing higher rates of temporary pacing (94.4% for ED+ vs. 28.9% for ED−, p < 0.001). Emergency LPM implantation was not correlated with an increase in the rate of major complications compared to the control group (6.9% ED+ vs. 4.2% ED−, p = 0.244). Conclusion. LPM implantation is a feasible procedure for the treatment of severe bradyarrhythmia in an urgent setting. Urgent LPM implantation was not correlated with an increase in the rate of major complications compared to the control group, but it was associated with longer procedural times.
No takes yet. Share an insight, caveat, or question.
Schiavone et al. (2022) conducted a cohort in bradyarrhythmia (n=1,154). Emergency leadless pacemaker (LPM) implantation vs. Elective LPM implantation was evaluated on peri-procedural complications (p=0.244). Emergency leadless pacemaker implantation was not associated with a significant increase in major peri-procedural complications compared to elective implantation (6.9% vs 4.2%, p=0.244).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: