Answers are grounded in this paper and cite their sources.
In this paper, a novel optical BioNEMS (Bio Nano-Electro-Mechanical-Systems) sensor using a nanomechanical plasmonic structure is proposed to develop a high performance label-free biosensor. The present device is composed of a functionalized BioNEMS cantilever coated by a gold layer as a part of the proposed plasmonic grating waveguide. This plasmonic system is based on the light interaction with a metallic nanostructure as a metal-insulator-metal (MIM) Bragg grating where an air gap acts as the insulator. Passing patient samples through the functionalized surface of the nanocantilever, the biological interactions release a stress on the surface because of binding events. The cantilever can be thus deflected by inducing a differential surface stress. Therefore, the air gap distance (the insulator width) between two parts of the plasmonic grating structure is changed to excite surface plasmon polaritons in a different wavelength. Furthermore, the present biosensor is analyzed by numerical and analytical approaches and its functional characteristics are obtained as follows: optical sensitivity of 7872 nm/RIU, figure of merit (FOM) of 394 RIU−1, mechanical sensitivity of 1.2 µm/Nm−1. According to the results, the proposed label-free identification method can be used as a sensitive platform for detection of biomarkers in point of care testing (POCT) applications.
No takes yet. Share an insight, caveat, or question.
Marvi et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: