PR logic algorithm in an ICD triggers inappropriate detection and shock for SVT.
Why the study?
Tachycardia detection algorithms in ICDs reduce but do not eliminate inappropriate shocks, and understanding algorithm limitations is important for clinical use.
Case Report (n=1)
The PR logic™ algorithm can misclassify SVT as dual tachycardia, leading to inappropriate ICD shocks.
Answers are grounded in this paper and cite their sources.
May prompt vigilance for inappropriate shocks with PR logic™ in SVT; leaves open need for algorithm validation in larger studies.
Tachycardia detection and therapy algorithms in Implantable Cardioverter-Defibrillators (ICD) reduce, but do not eliminate inappropriate ICD shocks. Awareness of the pros and cons of a particular algorithm helps to predict its utility in specific situations. We report a case where PR logic™, an algorithm commonly used in currently implanted ICDs to differentiate supraventricular tachycardia (SVT) from ventricular tachycardia resulted in inappropriate detection and shock for an SVT, and discuss several solutions to the problem.
No takes yet. Share an insight, caveat, or question.
Thachil et al. (2014) conducted a case report in Supraventricular tachycardia (n=1). PR logic™ algorithm was evaluated on Inappropriate detection and shock for SVT. The PR logic™ algorithm in an implantable cardioverter-defibrillator resulted in inappropriate detection and shock for a supraventricular tachycardia.